Titel: | Investigation of the imaging characteristics of the ALBIRA II small animal PET system for 18F, 68Ga and 64Cu |
Verf.angabe: | Ali Asgar Attarwala, Yvonne Wanjiku Karanja, Deni Hardiansyah, Chiara Romanó, Mareike Roscher, Björn Wängler, Gerhard Glatting |
E-Jahr: | 2017 |
Jahr: | June 2017 |
Umfang: | 13 S. |
Fussnoten: | Gesehen am 29.03.2018 |
Titel Quelle: | Enthalten in: Zeitschrift für medizinische Physik |
Ort Quelle: | Amsterdam [u.a.] : Elsevier, 1990 |
Jahr Quelle: | 2017 |
Band/Heft Quelle: | 27(2017), 2, Seite 132-144 |
ISSN Quelle: | 1876-4436 |
Abstract: | Aim: In this study the performance characteristics of the Albira II PET sub-system and the response of the system for the following radionuclides 18F, 68Ga and 64Cu was analyzed. Materials and methods: The Albira II tri-modal system (Bruker BioSpin MRI GmbH, Ettlingen, Germany) is a pre-clinical device for PET, SPECT and CT. The PET sub-system uses single continuous crystal detectors of lutetium yttrium orthosilicate (LYSO). The detector assembly consists of three rings of 8 detector modules. The transaxial field of view (FOV) has a diameter of 80mm and the axial FOV is 148mm. A NEMA NU-4 image quality phantom (Data Spectrum Corporation, Durham, USA) having five rods with diameters of 1, 2, 3, 4 and 5mm and a uniform central region was used. Measurements with 18F, 68Ga and 64Cu were performed in list mode acquisition over 10h. Data were reconstructed using a maximum-likelihood expectation-maximization (MLEM) algorithm with iteration numbers between 5 and 50. System sensitivity, count rate linearity, convergence and recovery coefficients were analyzed. Results: The sensitivities for the entire FOV (non-NEMA method) for 18F, 68Ga and 64Cu were (3.78±0.05)%, (3.97±0.18)% and (3.79±0.37)%, respectively. The sensitivity based on the NEMA protocol using the 22Na point source yielded (5.53±0.06)%. Dead-time corrected true counts were linear for activities ≤7MBq (18F and 68Ga) and ≤17MBq (64Cu) in the phantom. The radial, tangential and axial full widths at half maximum (FWHMs) were 1.52, 1.47 and 1.48mm. Recovery coefficients for the uniform region with a total activity of 8MBq in the phantom were (0.97±0.05), (0.98±0.06), (0.98±0.06) for 18F, 68Ga and 64Cu, respectively. Conclusion: The Albira II pre-clinical PET system has an adequate sensitivity range and the system linearity is suitable for the range of activities used for pre-clinical imaging. Overall, the system showed a favorable image quality for pre-clinical applications. |
DOI: | doi:10.1016/j.zemedi.2017.01.005 |
URL: | Bitte beachten Sie: Dies ist ein Bibliographieeintrag. Ein Volltextzugriff für Mitglieder der Universität besteht hier nur, falls für die entsprechende Zeitschrift/den entsprechenden Sammelband ein Abonnement besteht oder es sich um einen OpenAccess-Titel handelt.
Volltext: http://dx.doi.org/10.1016/j.zemedi.2017.01.005 |
| Volltext: http://www.sciencedirect.com/science/article/pii/S0939388917300247 |
| DOI: https://doi.org/10.1016/j.zemedi.2017.01.005 |
Datenträger: | Online-Ressource |
Sprache: | eng |
Sach-SW: | PET isotopes |
| PET-Isotope |
| Positron emission tomography (PET) |
| Positronen-Emissions-Tomographie (PET) |
| präklinische Bildgebung |
| pre-clinical imaging |
| recovery coefficients |
| Recovery-Koeffizienten |