Titel: | Ion recombination correction factor in scanned light-ion beams for absolute dose measurement using plane-parallel ionisation chambers |
Verf.angabe: | S. Rossomme, J. Horn, S. Brons, O. Jäkel, A. Mairani, M. Ciocca, V. Floquet, F. Romano, D. Rodriguez Garcia, S. Vynckier, H. Palmans |
Umfang: | 18 S. |
Fussnoten: | Gesehen am 25.04.2018 |
Titel Quelle: | Enthalten in: Physics in medicine and biology |
Jahr Quelle: | 2017 |
Band/Heft Quelle: | 62(2017), 13, S. 5365-5382 |
ISSN Quelle: | 1361-6560 |
Abstract: | Based on international reference dosimetry protocols for light-ion beams, a correction factor ( k s ) has to be applied to the response of a plane-parallel ionisation chamber, to account for recombination of negative and positive charges in its air cavity before these charges can be collected on the electrodes. In this work, k s for IBA PPC40 Roos-type chambers is investigated in four scanned light-ion beams (proton, helium, carbon and oxygen). To take into account the high dose-rates used with scanned beams and LET-values, experimental results are compared to a model combining two theories. One theory, developed by Jaffé, describes the variation of k s with the ionization density within the ion track (initial recombination) and the other theory, developed by Boag, describes the variation of k s with the dose rate (volume recombination). Excellent agreement is found between experimental and theoretical k s -values. All results confirm that k s cannot be neglected. The solution to minimise k s is to use the ionisation chamber at high voltage. However, one must be aware that charge multiplication may complicate the interpretation of the measurement. For the chamber tested, it was found that a voltage of 300 V can be used without further complication. As the initial recombination has a logarithmic variation as a function of 1/ V , the two-voltage method is not applicable to these scanned beams. |
DOI: | doi:10.1088/1361-6560/aa730f |
URL: | Bitte beachten Sie: Dies ist ein Bibliographieeintrag. Ein Volltextzugriff für Mitglieder der Universität besteht hier nur, falls für die entsprechende Zeitschrift/den entsprechenden Sammelband ein Abonnement besteht oder es sich um einen OpenAccess-Titel handelt.
Verlag: http://dx.doi.org/10.1088/1361-6560/aa730f |
| Verlag: http://stacks.iop.org/0031-9155/62/i=13/a=5365 |
| DOI: https://doi.org/10.1088/1361-6560/aa730f |
Datenträger: | Online-Ressource |
Sprache: | eng |
K10plus-PPN: | 1572361972 |
Verknüpfungen: | → Zeitschrift |
Ion recombination correction factor in scanned light-ion beams for absolute dose measurement using plane-parallel ionisation chambers / Rossomme, Severine [VerfasserIn] (Online-Ressource)