| Online-Ressource |
Verfasst von: | Maisary, Sameer al- [VerfasserIn]  |
| Graser, Bastian [VerfasserIn]  |
| Engelhardt, Sandy [VerfasserIn]  |
| Wolf, Ivo [VerfasserIn]  |
| Karck, Matthias [VerfasserIn]  |
| De Simone, Raffaele [VerfasserIn]  |
Titel: | The geometrical effect of different annuloplasty rings on mitral valve annulus |
Verf.angabe: | Sameer Al-Maisary, Bastian Graser, Sandy Engelhardt, Ivo Wolf, Matthias Karck, Raffaele DE Simone |
E-Jahr: | 2017 |
Jahr: | March 28, 2017 |
Umfang: | 8 S. |
Fussnoten: | Gesehen am 30.04.2018 |
Titel Quelle: | Enthalten in: The journal of cardiovascular surgery |
Ort Quelle: | Torino : Ed. Minerva Medica, 1998 |
Jahr Quelle: | 2017 |
Band/Heft Quelle: | 58(2017), 3, Seite 481-488 |
ISSN Quelle: | 1827-191X |
Abstract: | BACKGROUND: Different types of mitral annuloplasty rings are commercially available. The aim of this study was to investigate the effect of implantation of six types of annuloplasty rings on the geometry and dynamics of the mitral valve. METHODS: Three-dimensional echocardiography images of 42 patients were acquired to visualize the mitral valve annulus. Virtual representations of six commercially available annuloplasty rings were matched to anatomical mitral annuli of each patient according to anterolateral-posteromedial diameter. The virtual displacement of each annuloplasty ring after the implantation was measured and compared with the other rings. RESULTS: Patients with severe mitral regurgitation had significantly dilated annuli according to anterolateral-posteromedial diameter, anterior-posterior diameter and to annulus circumference. Anterior and posterior heights of the mitral annuli and non-planarity angle showed no significant differences among different patients with different degree of mitral regurgitation. The ratio of anterior-posterior to anterolateral-posteromedial diameter was almost identical in all groups with identical annular shapes. The implantation of the Carpentier-Edwards Classic Annuloplasty Ring™ led to maximal displacement of mitral annulus, followed by the IM-Ring™, without a statistical significance. In contrary, the implantation of a MyxoETlogix Ring™ was associated with minimal displacement of mitral annulus throughout the groups, but without statistical significance. CONCLUSIONS: The implantation of different ring types in patients with different annuli shapes and dimensions did not lead to any significant change in the configuration of mitral annuli after the virtual implantation of the tested annuloplasty rings. |
DOI: | doi:10.23736/S0021-9509.16.09314-9 |
URL: | Bitte beachten Sie: Dies ist ein Bibliographieeintrag. Ein Volltextzugriff für Mitglieder der Universität besteht hier nur, falls für die entsprechende Zeitschrift/den entsprechenden Sammelband ein Abonnement besteht oder es sich um einen OpenAccess-Titel handelt.
Volltext: http://dx.doi.org/10.23736/S0021-9509.16.09314-9 |
| DOI: https://doi.org/10.23736/S0021-9509.16.09314-9 |
Datenträger: | Online-Ressource |
Sprache: | eng |
Sach-SW: | Echocardiography, Three-Dimensional |
| Echocardiography, Transesophageal |
| Heart Valve Prosthesis |
| Heart Valve Prosthesis Implantation |
| Humans |
| Mitral Valve |
| Mitral Valve Annuloplasty |
| Mitral Valve Insufficiency |
| Prosthesis Design |
| Severity of Illness Index |
| Treatment Outcome |