Navigation überspringen
Universitätsbibliothek Heidelberg
Status: Bibliographieeintrag

Verfügbarkeit
Standort: ---
Exemplare: ---
heiBIB
 Online-Ressource
Verfasst von:Sommer, Christof-Matthias [VerfasserIn]   i
 Arnegger, Florian [VerfasserIn]   i
 Koch, Vitali [VerfasserIn]   i
 Pap, Bence [VerfasserIn]   i
 Holzschuh, Maria [VerfasserIn]   i
 Bellemann, Nadine [VerfasserIn]   i
 Gehrig, Tobias [VerfasserIn]   i
 Senft, Jonas [VerfasserIn]   i
 Nickel, Felix [VerfasserIn]   i
 Mogler, Carolin [VerfasserIn]   i
 Meinzer, Hans-Peter [VerfasserIn]   i
 Stampfl, Ulrike [VerfasserIn]   i
 Kauczor, Hans-Ulrich [VerfasserIn]   i
 Radeleff, Boris [VerfasserIn]   i
Titel:Microwave ablation of porcine kidneys in vivo
Titelzusatz:effect of two different ablation modes (“temperature control” and “power control”) on procedural outcome
Verf.angabe:C.M. Sommer, F. Arnegger, V. Koch, B. Pap, M. Holzschuh, N. Bellemann, T. Gehrig, J. Senft, F. Nickel, C. Mogler, S. Zelzer, H.P. Meinzer, U. Stampfl, H.U. Kauczor, B.A. Radeleff
Jahr:2012
Jahr des Originals:2011
Umfang:8 S.
Fussnoten:First Online: 12 May 2011 ; Gesehen am 24.07.2018
Titel Quelle:Enthalten in: CardioVascular and interventional radiology
Ort Quelle:Berlin : Springer, 1978
Jahr Quelle:2012
Band/Heft Quelle:35(2012), 3, Seite 653-660
ISSN Quelle:1432-086X
Abstract:Purpose: This study was designed to analyze the effect of two different ablation modes (“temperature control” and “power control”) of a microwave system on procedural outcome in porcine kidneys in vivo. Methods: A commercially available microwave system (Avecure Microwave Generator; MedWaves, San Diego, CA) was used. The system offers the possibility to ablate with two different ablation modes: temperature control and power control. Thirty-two microwave ablations were performed in 16 kidneys of 8 pigs. In each animal, one kidney was ablated twice by applying temperature control (ablation duration set point at 60 s, ablation temperature set point at 96°C, automatic power set point; group I). The other kidney was ablated twice by applying power control (ablation duration set point at 60 s, ablation temperature set point at 96°C, ablation power set point at 24 W; group II). Procedural outcome was analyzed: (1) technical success (e.g., system failures, duration of the ablation cycle), and (2) ablation geometry (e.g., long axis diameter, short axis diameter, and circularity). Results: System failures occurred in 0% in group I and 13% in group II. Duration of the ablation cycle was 60 ± 0 s in group I and 102 ± 21 s in group II. Long axis diameter was 20.3 ± 4.6 mm in group I and 19.8 ± 3.5 mm in group II (not significant (NS)). Short axis diameter was 10.3 ± 2 mm in group I and 10.5 ± 2.4 mm in group II (NS). Circularity was 0.5 ± 0.1 in group I and 0.5 ± 0.1 in group II (NS). Conclusions: Microwave ablations performed with temperature control showed fewer system failures and were finished faster. Both ablation modes demonstrated no significant differences with respect to ablation geometry.
DOI:doi:10.1007/s00270-011-0171-5
URL:Bitte beachten Sie: Dies ist ein Bibliographieeintrag. Ein Volltextzugriff für Mitglieder der Universität besteht hier nur, falls für die entsprechende Zeitschrift/den entsprechenden Sammelband ein Abonnement besteht oder es sich um einen OpenAccess-Titel handelt.

Volltext: http://dx.doi.org/10.1007/s00270-011-0171-5
 Volltext: https://link.springer.com/article/10.1007/s00270-011-0171-5
 DOI: https://doi.org/10.1007/s00270-011-0171-5
Datenträger:Online-Ressource
Sprache:eng
K10plus-PPN:157788602X
Verknüpfungen:→ Zeitschrift

Permanenter Link auf diesen Titel (bookmarkfähig):  https://katalog.ub.uni-heidelberg.de/titel/68288231   QR-Code
zum Seitenanfang