| Online-Ressource |
Verfasst von: | Sommer, Christof-Matthias [VerfasserIn]  |
| Arnegger, Florian [VerfasserIn]  |
| Koch, Vitali [VerfasserIn]  |
| Pap, Bence [VerfasserIn]  |
| Holzschuh, Maria [VerfasserIn]  |
| Bellemann, Nadine [VerfasserIn]  |
| Gehrig, Tobias [VerfasserIn]  |
| Senft, Jonas [VerfasserIn]  |
| Nickel, Felix [VerfasserIn]  |
| Mogler, Carolin [VerfasserIn]  |
| Meinzer, Hans-Peter [VerfasserIn]  |
| Stampfl, Ulrike [VerfasserIn]  |
| Kauczor, Hans-Ulrich [VerfasserIn]  |
| Radeleff, Boris [VerfasserIn]  |
Titel: | Microwave ablation of porcine kidneys in vivo |
Titelzusatz: | effect of two different ablation modes (“temperature control” and “power control”) on procedural outcome |
Verf.angabe: | C.M. Sommer, F. Arnegger, V. Koch, B. Pap, M. Holzschuh, N. Bellemann, T. Gehrig, J. Senft, F. Nickel, C. Mogler, S. Zelzer, H.P. Meinzer, U. Stampfl, H.U. Kauczor, B.A. Radeleff |
Jahr: | 2012 |
Jahr des Originals: | 2011 |
Umfang: | 8 S. |
Fussnoten: | First Online: 12 May 2011 ; Gesehen am 24.07.2018 |
Titel Quelle: | Enthalten in: CardioVascular and interventional radiology |
Ort Quelle: | Berlin : Springer, 1978 |
Jahr Quelle: | 2012 |
Band/Heft Quelle: | 35(2012), 3, Seite 653-660 |
ISSN Quelle: | 1432-086X |
Abstract: | Purpose: This study was designed to analyze the effect of two different ablation modes (“temperature control” and “power control”) of a microwave system on procedural outcome in porcine kidneys in vivo. Methods: A commercially available microwave system (Avecure Microwave Generator; MedWaves, San Diego, CA) was used. The system offers the possibility to ablate with two different ablation modes: temperature control and power control. Thirty-two microwave ablations were performed in 16 kidneys of 8 pigs. In each animal, one kidney was ablated twice by applying temperature control (ablation duration set point at 60 s, ablation temperature set point at 96°C, automatic power set point; group I). The other kidney was ablated twice by applying power control (ablation duration set point at 60 s, ablation temperature set point at 96°C, ablation power set point at 24 W; group II). Procedural outcome was analyzed: (1) technical success (e.g., system failures, duration of the ablation cycle), and (2) ablation geometry (e.g., long axis diameter, short axis diameter, and circularity). Results: System failures occurred in 0% in group I and 13% in group II. Duration of the ablation cycle was 60 ± 0 s in group I and 102 ± 21 s in group II. Long axis diameter was 20.3 ± 4.6 mm in group I and 19.8 ± 3.5 mm in group II (not significant (NS)). Short axis diameter was 10.3 ± 2 mm in group I and 10.5 ± 2.4 mm in group II (NS). Circularity was 0.5 ± 0.1 in group I and 0.5 ± 0.1 in group II (NS). Conclusions: Microwave ablations performed with temperature control showed fewer system failures and were finished faster. Both ablation modes demonstrated no significant differences with respect to ablation geometry. |
DOI: | doi:10.1007/s00270-011-0171-5 |
URL: | Bitte beachten Sie: Dies ist ein Bibliographieeintrag. Ein Volltextzugriff für Mitglieder der Universität besteht hier nur, falls für die entsprechende Zeitschrift/den entsprechenden Sammelband ein Abonnement besteht oder es sich um einen OpenAccess-Titel handelt.
Volltext: http://dx.doi.org/10.1007/s00270-011-0171-5 |
| Volltext: https://link.springer.com/article/10.1007/s00270-011-0171-5 |
| DOI: https://doi.org/10.1007/s00270-011-0171-5 |
Datenträger: | Online-Ressource |
Sprache: | eng |
K10plus-PPN: | 157788602X |
Verknüpfungen: | → Zeitschrift |
Microwave ablation of porcine kidneys in vivo / Sommer, Christof-Matthias [VerfasserIn]; 2012 (Online-Ressource)