Titel:Wire Placement in the Sustentaculum Tali Using a 2D Projection-Based Software Application for Mobile C-Arms
Titelzusatz:Cadaveric Study
Verf.angabe:Benedict Swartman, Dirk Frere, Wei Wei, Marc Schnetzke, Stephan Grechenig, Amir Matityahu, Nils Beisemann, Holger Keil, Jochen Franke, Paul Alfred Grützner, and Sven Yves Vetter
E-Jahr:2018
Jahr:January 19, 2018
Umfang:8 S.
Fussnoten:Gesehen am 30.04.2019
Titel Quelle:Enthalten in: Foot & ankle international
Ort Quelle:Thousand Oaks, Calif. : Sage, 1980
Jahr Quelle:2018
Band/Heft Quelle:39(2018), 4, Seite 485-492
ISSN Quelle:1944-7876
Abstract:Background:Indirect screw fixation of the sustentaculum tali in the lateral-medial direction can be challenging due to the complex calcaneal anatomy. A novel 2-dimensional (2D) projection-based software application detects Kirschner wires (K-wires) and visualizes their intended direction as a colored trajectory. The aim of this prospectively randomized cadaver study was to investigate whether the software would facilitate the indirect K-wire placement in the sustentaculum tali.Methods:In 20 cadaver foot specimens, K-wires were placed indirectly in the sustentaculum tali by an experienced and an inexperienced surgeon, with and without using the application. Number of placement attempts, duration of procedure, fluoroscopy time, and number of individual fluoroscopy images were recorded. Each wire?s position was analyzed in a 3-dimensional (3D) C-arm scan by an experienced blinded investigator.Results:Use of the software by the inexperienced surgeon significantly reduced the number of placement attempts from 3.2 to 1.2 (P = .006). The application also reduced operating time, from 273 s to 199 s (P = .15), and fluoroscopy time, from 41 s to 29 s (P = .15). Using the software, the experienced surgeon had a longer operating time (139 s to 183 s; P = .30), longer fluoroscopy time (5.6 s to 9.2 s; P = .17), and more individual fluoroscopy images (11.6 to 14.8; P = .30). Wire position did not show significant differences in both cases.Conclusion:During indirect K-wire placement in the sustentaculum tali, the software appeared to be a useful tool for the inexperienced surgeon. In our chosen study setting, the experienced surgeon did not benefit from the software.Clinical Relevance:Possible indications for the software would be fractures of the proximal femur, sacrum, sacroiliac instabilities, vertebral bodies, scaphoid, Lisfranc joint, talus and calcaneus.
DOI:doi:10.1177/1071100717746618
URL:Bitte beachten Sie: Dies ist ein Bibliographieeintrag. Ein Volltextzugriff für Mitglieder der Universität besteht hier nur, falls für die entsprechende Zeitschrift/den entsprechenden Sammelband ein Abonnement besteht oder es sich um einen OpenAccess-Titel handelt.

Volltext: https://doi.org/10.1177/1071100717746618
 Volltext: https://doi.org/10.1177/1071100717746618
 DOI: https://doi.org/10.1177/1071100717746618
Datenträger:Online-Ressource
Sprache:eng
K10plus-PPN:166381192X
Verknüpfungen:→ Zeitschrift

Permanenter Link auf diesen Titel (bookmarkfähig):  https://katalog.ub.uni-heidelberg.de/titel/68383702   QR-Code
zum Seitenanfang