Titel: | Patient-specific online dose verification based on transmission detector measurements |
Verf.angabe: | Johannes Thoelking, Jens Fleckenstein, Yuvaraj Sekar, Ramesh Boggula, Frank Lohr, Frederik Wenz, Hansjoerg Wertz |
E-Jahr: | 2016 |
Jahr: | 16 April 2016 |
Umfang: | 6 S. |
Fussnoten: | Gesehen am 20.05.2019 |
Titel Quelle: | Enthalten in: Radiotherapy and oncology |
Ort Quelle: | Amsterdam [u.a.] : Elsevier Science, 1983 |
Jahr Quelle: | 2016 |
Band/Heft Quelle: | 119(2016), 2, Seite 351-356 |
ISSN Quelle: | 1879-0887 |
Abstract: | Background and purpose - Since IMRT-techniques lead to an increasingly complicated environment, a patient specific IMRT-plan verification is recommended. Furthermore, verifications during patient irradiation and 3D dose reconstruction have the potential to improve treatment delivery, accuracy and safety. This study provides a detailed investigation of the new transmission detector (DTD) Dolphin (IBA Dosimetry, Germany) for online dosimetry. - Materials and methods - The clinical performance of the DTD was tested by dosimetric plan verification in 2D and 3D for 18 IMRT-sequences. In 2D, DTD measurements were compared to a pre-treatment verification method and a treatment planning system by gamma index and dose difference evaluations. In 3D, dose-volume-histogram (DVH) indices and gamma analysis were evaluated. Furthermore, the error detection ability was tested with leaf position uncertainties and deviations in the linear accelerator (LINAC) output. - Results - The DTD measurements were in excellent agreement to reference measurements in both 2D (γ3%,3mm=(99.7±0.6)% <1, ΔD±5%=(99.5±0.5)%) and 3D. Only a small dose underestimation (<2%) within the target volume was observed when analyzing DVH-indices. Positional errors of the leaf banks larger than 1mm and errors in LINAC output larger than 2% were identified with the DTD. - Conclusions - The DTD measures the delivered dose with sufficient accuracy and is therefore suitable for clinical routine. |
DOI: | doi:10.1016/j.radonc.2016.04.003 |
URL: | Bitte beachten Sie: Dies ist ein Bibliographieeintrag. Ein Volltextzugriff für Mitglieder der Universität besteht hier nur, falls für die entsprechende Zeitschrift/den entsprechenden Sammelband ein Abonnement besteht oder es sich um einen OpenAccess-Titel handelt.
Volltext: https://doi.org/10.1016/j.radonc.2016.04.003 |
| Volltext: http://www.sciencedirect.com/science/article/pii/S0167814016310374 |
| DOI: https://doi.org/10.1016/j.radonc.2016.04.003 |
Datenträger: | Online-Ressource |
Sprache: | eng |
Sach-SW: | Online dose verification |
| Quality assurance |
| Transmission detector |