| Online-Ressource |
Verfasst von: | Emmerich, Julian [VerfasserIn] ![i](/opacicon/information2.png) |
| Laun, Frederik B. [VerfasserIn] ![i](/opacicon/information2.png) |
| Pfaffenberger, Asja [VerfasserIn] ![i](/opacicon/information2.png) |
| Schilling, Rebecca Anna [VerfasserIn] ![i](/opacicon/information2.png) |
| Denoix, Michael [VerfasserIn] ![i](/opacicon/information2.png) |
| Maier, Florian [VerfasserIn] ![i](/opacicon/information2.png) |
| Sterzing, Florian [VerfasserIn] ![i](/opacicon/information2.png) |
| Bostel, Tilman [VerfasserIn] ![i](/opacicon/information2.png) |
| Straub, Sina [VerfasserIn] ![i](/opacicon/information2.png) |
Titel: | Technical note |
Titelzusatz: | on the size of susceptibility-induced MR image distortions in prostate and cervix in the context of MR-guided radiation therapy |
Verf.angabe: | Julian Emmerich, Frederik B. Laun, Asja Pfaffenberger, Rebecca Schilling, Michael Denoix, Florian Maier, Florian Sterzing, Tilman Bostel, Sina Straub |
E-Jahr: | 2018 |
Jahr: | 02 February 2018 |
Umfang: | 8 S. |
Fussnoten: | Gesehen am 23.10.2019 |
Titel Quelle: | Enthalten in: Medical physics |
Ort Quelle: | Hoboken, NJ : Wiley, 1974 |
Jahr Quelle: | 2018 |
Band/Heft Quelle: | 45(2018), 4, Seite 1586-1593 |
ISSN Quelle: | 2473-4209 |
| 1522-8541 |
Abstract: | Purpose To investigate the extent of MR image distortions in the pelvis caused by susceptibility-induced field inhomogeneities in MR images in the context of a study on MR-guided radiotherapy. Methods Using a high-bandwidth double-echo gradient echo sequence, field maps and distortion maps of the pelvis were calculated and evaluated for 219 exams (92 of female and 127 of male patients) to investigate patient-related image distortions caused by susceptibility differences in an ongoing study on MR-guided radiotherapy. The evaluation of distortions in the regions “rectum”, “prostate”, “cervix”, and in a reference region in the gluteus maximus was based on masks drawn by two readers. Results Distortions in the prostate and cervix were smaller than 0.03 px (0.1 mm) for 99% of voxels, and reached a maximum value of 0.09 px (0.3 mm). In the reference region, maximum distortions were smaller than in the prostate and cervix. Conclusions Using a geometric uncertainty of 0.2 px (0.6 mm) in margin definition for organs that are close to the rectum like the prostate and the cervix would be a cautious choice to account for susceptibility-induced distortions that can arise during MR-based treatment guidance for the imaging setting used in this study. Since distortions are inversely proportional to the readout bandwidth of the sequence, safety margins need to be adapted adequately. Additional sources of image distortions like gradient nonlinearities are not included in our margin recommendations and should be considered separately. |
DOI: | doi:10.1002/mp.12785 |
URL: | Bitte beachten Sie: Dies ist ein Bibliographieeintrag. Ein Volltextzugriff für Mitglieder der Universität besteht hier nur, falls für die entsprechende Zeitschrift/den entsprechenden Sammelband ein Abonnement besteht oder es sich um einen OpenAccess-Titel handelt.
Volltext: https://doi.org/10.1002/mp.12785 |
| Verlag: https://aapm.onlinelibrary.wiley.com/doi/abs/10.1002/mp.12785 |
| DOI: https://doi.org/10.1002/mp.12785 |
Datenträger: | Online-Ressource |
Sprache: | eng |
Sach-SW: | distortions |
| geometric accuracy |
| IGRT |
| MR-guided radiotherapy |
| MRI |
| safety margins |
K10plus-PPN: | 1679460161 |
Verknüpfungen: | → Zeitschrift |
Technical note / Emmerich, Julian [VerfasserIn]; 02 February 2018 (Online-Ressource)