Navigation überspringen
Universitätsbibliothek Heidelberg
Status: Bibliographieeintrag

Verfügbarkeit
Standort: ---
Exemplare: ---
heiBIB
 Online-Ressource
Verfasst von:Emmerich, Julian [VerfasserIn]   i
 Laun, Frederik B. [VerfasserIn]   i
 Pfaffenberger, Asja [VerfasserIn]   i
 Schilling, Rebecca Anna [VerfasserIn]   i
 Denoix, Michael [VerfasserIn]   i
 Maier, Florian [VerfasserIn]   i
 Sterzing, Florian [VerfasserIn]   i
 Bostel, Tilman [VerfasserIn]   i
 Straub, Sina [VerfasserIn]   i
Titel:Technical note
Titelzusatz:on the size of susceptibility-induced MR image distortions in prostate and cervix in the context of MR-guided radiation therapy
Verf.angabe:Julian Emmerich, Frederik B. Laun, Asja Pfaffenberger, Rebecca Schilling, Michael Denoix, Florian Maier, Florian Sterzing, Tilman Bostel, Sina Straub
E-Jahr:2018
Jahr:02 February 2018
Umfang:8 S.
Fussnoten:Gesehen am 23.10.2019
Titel Quelle:Enthalten in: Medical physics
Ort Quelle:Hoboken, NJ : Wiley, 1974
Jahr Quelle:2018
Band/Heft Quelle:45(2018), 4, Seite 1586-1593
ISSN Quelle:2473-4209
 1522-8541
Abstract:Purpose To investigate the extent of MR image distortions in the pelvis caused by susceptibility-induced field inhomogeneities in MR images in the context of a study on MR-guided radiotherapy. Methods Using a high-bandwidth double-echo gradient echo sequence, field maps and distortion maps of the pelvis were calculated and evaluated for 219 exams (92 of female and 127 of male patients) to investigate patient-related image distortions caused by susceptibility differences in an ongoing study on MR-guided radiotherapy. The evaluation of distortions in the regions “rectum”, “prostate”, “cervix”, and in a reference region in the gluteus maximus was based on masks drawn by two readers. Results Distortions in the prostate and cervix were smaller than 0.03 px (0.1 mm) for 99% of voxels, and reached a maximum value of 0.09 px (0.3 mm). In the reference region, maximum distortions were smaller than in the prostate and cervix. Conclusions Using a geometric uncertainty of 0.2 px (0.6 mm) in margin definition for organs that are close to the rectum like the prostate and the cervix would be a cautious choice to account for susceptibility-induced distortions that can arise during MR-based treatment guidance for the imaging setting used in this study. Since distortions are inversely proportional to the readout bandwidth of the sequence, safety margins need to be adapted adequately. Additional sources of image distortions like gradient nonlinearities are not included in our margin recommendations and should be considered separately.
DOI:doi:10.1002/mp.12785
URL:Bitte beachten Sie: Dies ist ein Bibliographieeintrag. Ein Volltextzugriff für Mitglieder der Universität besteht hier nur, falls für die entsprechende Zeitschrift/den entsprechenden Sammelband ein Abonnement besteht oder es sich um einen OpenAccess-Titel handelt.

Volltext: https://doi.org/10.1002/mp.12785
 Verlag: https://aapm.onlinelibrary.wiley.com/doi/abs/10.1002/mp.12785
 DOI: https://doi.org/10.1002/mp.12785
Datenträger:Online-Ressource
Sprache:eng
Sach-SW:distortions
 geometric accuracy
 IGRT
 MR-guided radiotherapy
 MRI
 safety margins
K10plus-PPN:1679460161
Verknüpfungen:→ Zeitschrift

Permanenter Link auf diesen Titel (bookmarkfähig):  https://katalog.ub.uni-heidelberg.de/titel/68444273   QR-Code
zum Seitenanfang