| Online-Ressource |
Verfasst von: | El-Battrawy, Ibrahim [VerfasserIn]  |
| Besler, Johanna [VerfasserIn]  |
| Ansari, Uzair [VerfasserIn]  |
| Liebe, Volker [VerfasserIn]  |
| Schimpf, Rainer [VerfasserIn]  |
| Tueluemen, Erol [VerfasserIn]  |
| Rudic, Boris [VerfasserIn]  |
| Lang, Siegfried [VerfasserIn]  |
| Odening, Katja [VerfasserIn]  |
| Cyganek, Lukas [VerfasserIn]  |
| Wolpert, Christian [VerfasserIn]  |
| Zhou, Xiao-Bo [VerfasserIn]  |
| Borggrefe, Martin [VerfasserIn]  |
| Akın, Ibrahim [VerfasserIn]  |
Titel: | Long-term follow-up of implantable cardioverter-defibrillators in Short QT syndrome |
Verf.angabe: | Ibrahim El-Battrawy, Johanna Besler, Uzair Ansari, Volker Liebe, Rainer Schimpf, Erol Tülümen, Boris Rudic, Siegfried Lang, Katja Odening, Lukas Cyganek, Christian Wolpert, Xiaobo Zhou, Martin Borggrefe, Ibrahim Akin |
E-Jahr: | 2019 |
Jahr: | 16 March 2019 |
Umfang: | 7 S. |
Fussnoten: | Gesehen am 10.12.2019 |
Titel Quelle: | Enthalten in: Clinical research in cardiology |
Ort Quelle: | Berlin : Springer, 2006 |
Jahr Quelle: | 2019 |
Band/Heft Quelle: | 108(2019), 10, Seite 1140-1146 |
ISSN Quelle: | 1861-0692 |
Abstract: | BACKGROUND: Short QT syndrome (SQTS) is associated with sudden cardiac death and implantable cardioverter-defibrillator (ICD) implantation is recommended in this rare disease. However, only a few SQTS families have been reported in literature with limited follow-up data. - OBJECTIVES: In the recent study, we describe the outcome data of 57 SQTS patients receiving ICD implantation. This includes seven SQTS families consecutively admitted to our hospital between 2002 and 2017 as well as patients reported in published literature. - METHODS: Seven SQTS patients admitted to our hospital were followed up. Additionally, 7 studies out of a total of 626 researched articles were identified through systematic database search (PubMed, Web of Science, Cochrane Library, and Cinahl) and their data analyzed according to our model. - RESULTS: Complications during a median follow-up time of 67.4 months (IQR 6-162 months) were documented in 31 (54%) patients. Inappropriate shocks were seen in 33% due to T wave oversensing (8.7%), supraventricular tachycardia (19%), lead failure and fracture (21%). Further complications were infection (10%), battery depletion (7%) and psychological distress (3.5%). Appropriate shocks were documented in 19%. Three patients (5%) were treated with s-ICD due to recurrent complications of transvenous ICD. - CONCLUSION: ICD therapy is an effective therapy in SQTS patients. However, it is also associated with significant risk of device-related complications. |
DOI: | doi:10.1007/s00392-019-01449-3 |
URL: | Bitte beachten Sie: Dies ist ein Bibliographieeintrag. Ein Volltextzugriff für Mitglieder der Universität besteht hier nur, falls für die entsprechende Zeitschrift/den entsprechenden Sammelband ein Abonnement besteht oder es sich um einen OpenAccess-Titel handelt.
Volltext: http://dx.doi.org/10.1007/s00392-019-01449-3 |
| DOI: https://doi.org/10.1007/s00392-019-01449-3 |
Datenträger: | Online-Ressource |
Sprache: | eng |
Sach-SW: | ICD-related complications |
| Short QT syndrome |
| Sudden cardiac death |
K10plus-PPN: | 1684935970 |
Verknüpfungen: | → Zeitschrift |
Long-term follow-up of implantable cardioverter-defibrillators in Short QT syndrome / El-Battrawy, Ibrahim [VerfasserIn]; 16 March 2019 (Online-Ressource)