Titel:The Neuroform Atlas stent to assist coil embolization of intracranial aneurysms
Titelzusatz:a multicentre experience
Verf.angabe:Christian Ulfert, Mirko Pham, Michael Sonnberger, Franco Amaya, Johannes Trenkler, Martin Bendszus, Markus A. Möhlenbruch
E-Jahr:2018
Jahr:20 April 2018
Umfang:5 S.
Fussnoten:Gesehen am 19.02.2020
Titel Quelle:Enthalten in: Journal of neuroInterventional surgery
Ort Quelle:London : BMJ Journals, 2009
Jahr Quelle:2018
Band/Heft Quelle:10(2018), 12, Seite 1192-1196
ISSN Quelle:1759-8486
Abstract:Objective To assess the clinical safety and efficacy of the Atlas microstent in stent-assisted coil embolization of wide-necked intracranial aneurysms. - Methods Single-center observational study in 36 patients (24 female, 12 male, mean age 56 years) with 37 aneurysms for the endovascular treatment of wide-necked aneurysms. After giving informed consent, patients were included according to the following criteria: aneurysm dome-to-neck ratio <2 or neck diameter >4 mm, and a parent vessel diameter of ≤4.5 mm. Primary endpoint for clinical safety was absence of death, absence of major or minor stroke, and absence of transient ischemic attack. Primary endpoint for treatment efficacy was complete angiographic occlusion according to the Raymond-Roy occlusion classification (RROC) immediately after the procedure. - Results In 36/37 (97%) cases, the primary endpoint of safety was reached, one patient had a transitory ischemic attack which completely resolved until discharge. In 31/37 (84%) cases, complete occlusion (RROC 1) was reached, and in 6/36 (17%), a residual neck remained (RROC 2). A sequential approach (first stent, then coiling through the same catheter) was used in 21 cases; the other 16 were treated with the jailing technique. Deployment was technically successful in all cases. Follow-up at a median of 6.1 months was available for 29/37 (78%) aneurysms and showed complete occlusion in 27/29 aneurysms (93%) and a neck remnant in 2 cases (7%). - Conclusion Deployment of the Neuroform Atlas microstent is a safe and effective method for the treatment of intracranial wide-necked aneurysms.
DOI:doi:10.1136/neurintsurg-2017-013516
URL:Bitte beachten Sie: Dies ist ein Bibliographieeintrag. Ein Volltextzugriff für Mitglieder der Universität besteht hier nur, falls für die entsprechende Zeitschrift/den entsprechenden Sammelband ein Abonnement besteht oder es sich um einen OpenAccess-Titel handelt.

Volltext ; Verlag: https://doi.org/10.1136/neurintsurg-2017-013516
 Volltext: https://jnis.bmj.com/content/10/12/1192
 DOI: https://doi.org/10.1136/neurintsurg-2017-013516
Datenträger:Online-Ressource
Sprache:eng
Sach-SW:aneurysm
 angiography
 coil
 stent
K10plus-PPN:169037960X
Verknüpfungen:→ Zeitschrift

Permanenter Link auf diesen Titel (bookmarkfähig):  https://katalog.ub.uni-heidelberg.de/titel/68491173   QR-Code
zum Seitenanfang