Titel:Tomotherapy radiosurgery for arteriovenous malformations
Titelzusatz:current possibilities and future options with helical tomotherapy dynamic jaws?
Verf.angabe:S. Krause, M.D., S. Beck, M.D., O. Schramm, Ph.D., K. Schubert, Ph.D., H. Hauswald, M.D., A. Zabel-du Bois, M.D., K. Herfarth, M.D., J. Debus, M.D., Ph.D., F. Sterzing, M.D.
E-Jahr:2013
Jahr:October 1, 2013
Umfang:8 S.
Fussnoten:Gesehen am 28.10.2021
Titel Quelle:Enthalten in: Technology in cancer research & treatment
Ort Quelle:Thousand Oaks, CA : Sage Publishing, 2002
Jahr Quelle:2013
Band/Heft Quelle:12(2013), 5, Seite 421-428
ISSN Quelle:1533-0338
Abstract:This planning study was performed to compare stereotactic linac based radiosurgery of Arteriovenous Malformations (AVM) with current Helical Tomotherapy (HT) and future HT techniques. For 10 patients with AVM, dose distributions and treatment times of “regular” HT delivery (Reg 2.5/1/0.6 cm field width), Running-Start-Stop Treatment (RSS 5/2.5 cm), Axial Mode (Axial 5 cm) and Dynamic Jaw/Dynamic Couch delivery with a maximum field width of 5 cm (DJDC 5) were analysed and compared to linac-based stereotactic radiosurgery. Axial produced the fastest treatment (Axial 4:47 min vs. Linac 32:42 min) at the cost of large brain exposure (V10% 289 ml). Except for Reg 0.6, all other HT techniques achieved significantly shorter treatment times than linac-based treatment (e.g. Reg 1, 19:42 min, DJDC 6:30 min). However, high-dose brain exposure (V60%) was higher in all HT plans (e.g. Reg 0.6, 10 ml, Linac 9 ml), and only Reg 0.6 showed better low-dose exposure (V10% of 167 ml vs. 199 ml, not significant). Neither current nor future HT modes in their current version outperformed linac-based stereotactic radiosurgery. However, AVM with special geometry might still benefit from HT.
DOI:doi:10.7785/tcrt.2012.500335
URL:Bitte beachten Sie: Dies ist ein Bibliographieeintrag. Ein Volltextzugriff für Mitglieder der Universität besteht hier nur, falls für die entsprechende Zeitschrift/den entsprechenden Sammelband ein Abonnement besteht oder es sich um einen OpenAccess-Titel handelt.

Volltext: https://doi.org/10.7785/tcrt.2012.500335
 DOI: https://doi.org/10.7785/tcrt.2012.500335
Datenträger:Online-Ressource
Sprache:eng
Sach-SW:Arteriovenous malformation
 Dynamic jaws/Dynamic couch
 Helical tomotherapy
 Radiosurgery
K10plus-PPN:1752686454
Verknüpfungen:→ Zeitschrift

Permanenter Link auf diesen Titel (bookmarkfähig):  https://katalog.ub.uni-heidelberg.de/titel/68717778   QR-Code
zum Seitenanfang