Verfasst von:Heitzmann, Daniel [VerfasserIn]   i
 Günther, Michael [VerfasserIn]   i
 Becher, Benjamin [VerfasserIn]   i
 Alimusaj, Merkur [VerfasserIn]   i
 Block, Julia [VerfasserIn]   i
 Drongelen, Stefan van [VerfasserIn]   i
 Dreher, Thomas [VerfasserIn]   i
 Braatz, Frank [VerfasserIn]   i
 Wolf, Sebastian Immanuel [VerfasserIn]   i
Titel:Integrating strength tests of amputees within the protocol of conventional clinical gait analysis
Titelzusatz:a novel approach
Verf.angabe:Daniel W. W. Heitzmann, Michael Guenther, Benjamin Becher, Merkur Alimusaj, Julia Block, Stefan van Drongelen, Thomas Dreher, Frank Braatz and Sebastian I. Wolf
E-Jahr:2013
Jahr:2. März 2013
Umfang:10 S.
Teil:volume:58
 year:2013
 number:2
 pages:195-204
 extent:10
Fussnoten:Gesehen am 29.03.2021
Titel Quelle:Enthalten in: Biomedical engineering
Ort Quelle:Berlin [u.a.] : de Gruyter, 1998
Jahr Quelle:2013
Band/Heft Quelle:58(2013), 2, Seite 195-204
ISSN Quelle:1862-278X
Abstract:Clinical experience tells us that the lower-limb amputees are one of the patient groups who clearly suffer from a strength deficit in their involved side. However, there is no obvious evidence for the relation between the residual limb strength and walking ability in this population. Correlating the results of the conventional clinical gait analysis (CGA) with strength tests could help to find out how deficits in strength impact the amputees’ gait. In this contribution, a new device for measuring the isometric muscle strength of the hip and the knee was tested for feasibility. Three groups were tested: one group of 11 healthy subjects (29±5 years) to test the repeatability of the device, two unilateral amputees (one transfemoral for 56 years, one transtibial for 65 years), and a reference group of 17 healthy subjects (55±10 years). The new method presents an adequate technique to integrate strength testing within a standard protocol of the CGA. Results showed to be repeatable within sessions [i.e., within-day, intraclass correlation coefficient (ICC)>0.972] and between repeated measurements (i.e., day-to-day, ICC>0.765). The tested amputees showed clear deficits in maximum isometric joint moments in their most distal joint. The first results suggest evidence for a relation between the maximum isometric joint moments and gait deviations in amputees.
DOI:doi:10.1515/bmt-2012-0036
URL:Bitte beachten Sie: Dies ist ein Bibliographieeintrag. Ein Volltextzugriff für Mitglieder der Universität besteht hier nur, falls für die entsprechende Zeitschrift/den entsprechenden Sammelband ein Abonnement besteht oder es sich um einen OpenAccess-Titel handelt.

Volltext ; Verlag: https://doi.org/10.1515/bmt-2012-0036
 Volltext: https://www.degruyter.com/document/doi/10.1515/bmt-2012-0036/html
 DOI: https://doi.org/10.1515/bmt-2012-0036
Datenträger:Online-Ressource
Sprache:eng
K10plus-PPN:1752701232
Verknüpfungen:→ Zeitschrift

Permanenter Link auf diesen Titel (bookmarkfähig):  https://katalog.ub.uni-heidelberg.de/titel/68717839   QR-Code
zum Seitenanfang