Verfasst von:Roser, Christoph [VerfasserIn]   i
 Hodecker, Lutz [VerfasserIn]   i
 Köbel, Chris [VerfasserIn]   i
 Lux, Christopher J. [VerfasserIn]   i
 Ruckes, Dorothee [VerfasserIn]   i
 Rues, Stefan [VerfasserIn]   i
 Zenthöfer, Andreas [VerfasserIn]   i
Titel:Mechanical properties of CAD/CAM-fabricated in comparison to conventionally fabricated functional regulator 3 appliances
Verf.angabe:Christoph Roser, Lutz D. Hodecker, Chris Koebel, Christopher J. Lux, Dorothee Ruckes, Stefan Rues and Andreas Zenthöfer
E-Jahr:2021
Jahr:19 July 2021
Umfang:10 S.
Fussnoten:Gesehen am 24.09.2021
Titel Quelle:Enthalten in: Scientific reports
Ort Quelle:[London] : Macmillan Publishers Limited, part of Springer Nature, 2011
Jahr Quelle:2021
Band/Heft Quelle:11(2021), Artikel-ID 14719, Seite 1-10
ISSN Quelle:2045-2322
Abstract:Manufacturing of Fränkel's functional regulator 3 (FR3) is complicated and requires extensive knowledge from the dental technician. To determine whether FR3s produced by CAD/CAM techniques (CAD-FR3) meet similar mechanical properties like conventional FR3s (Con-FR3), for each of 10 patient cases, three CAD-FR3 designs (palatal connector cross-section 3 × 3 mm, 4 × 1 mm or 5 × 2 mm) and one Con-FR3 were subjected to cyclic loading and subsequent fracture testing in a universal testing device. Transversal load capacity (Fmax(FR3)) and stiffness were compared among the different CAD-FR3 designs and Con-FR3s using Friedman and Wilcoxon tests with a significance level of α = 0.05. All CAD-FR3 designs had significantly higher mean Fmax(FR3) (p ≤ 0.007) and stiffness (p ≤ 0.005) than the Con-FR3s. The CAD-FR33×3 had the highest mean Fmax(FR3) (98.2 ± 26.2 N) and stiffness (37.1 ± 15.5 N/mm), closely followed by the CAD-FR35×2 (Fmax(FR3): 90.3 ± 24.7 N; stiffness: 30.0 ± 12.3 N/mm). Among the CAD appliances, CAD-FR34×1 had the lowest values (p ≤ 0.007 for all pairwise tests) with Fmax(FR3) of 45.8 ± 17.9 N and stiffness of 12.5 ± 7.3 N/mm. CAD-FR3s have superior mechanical properties in comparison to Con-FR3s if certain design parameters are followed. Further clinical investigations have to examine if they can serve as an alternative in practice.
DOI:doi:10.1038/s41598-021-94237-x
URL:kostenfrei: Volltext: https://doi.org/10.1038/s41598-021-94237-x
 kostenfrei: Volltext: https://www.nature.com/articles/s41598-021-94237-x
 DOI: https://doi.org/10.1038/s41598-021-94237-x
Datenträger:Online-Ressource
Sprache:eng
K10plus-PPN:1771774789
Verknüpfungen:→ Zeitschrift
 
 
Lokale URL UB: Zum Volltext

Permanenter Link auf diesen Titel (bookmarkfähig):  https://katalog.ub.uni-heidelberg.de/titel/68783146   QR-Code
zum Seitenanfang