Titel:Initial experience with the Trevo NXT stent retriever
Verf.angabe:Manina M. Etter, Markus Möhlenbruch, Charlotte S. Weyland, Carlos Pérez-García, Manuel Moreu, Francesco Capasso, Nicola Limbucci, Omid Nikoubashman, Martin Wiesmann, Kristine Blackham, Ioannis Tsogkas, Peter Sporns, Johanna Maria Ospel, Alex Brehm and Marios-Nikos Psychogios
E-Jahr:2021
Jahr:16 July 2021
Umfang:6 S.
Teil:volume:12
 year:2021
 month:07
 elocationid:704329
 pages:1-6
 extent:6
Fussnoten:Gesehen am 06.10.2021
Titel Quelle:Enthalten in: Frontiers in neurology
Ort Quelle:Lausanne : Frontiers Research Foundation, 2008
Jahr Quelle:2021
Band/Heft Quelle:12(2021) vom: Juli, Artikel-ID 704329, Seite 1-6
ISSN Quelle:1664-2295
Abstract:Background: The application of a new coating to the delivery wire of the Trevo retriever has the potential to improve its handling. We therefore report our initial experience with this new stent retriever for mechanical thrombectomy of large and medium vessel occlusions.Methods: We pooled data of four high-volume European stroke centers over the time period from October 2020 to February 2021. Patients were included in our study if the Trevo NXT stent retriever was used as a first-line device. Primary endpoints were first-pass near-complete or complete reperfusion, defined as mTICI score of ≥2c. Secondary endpoints were final reperfusion, National Institutes of Health Stroke Scale (NIHSS) at 24 h and discharge, device malfunctions, complications during the procedure, and subjective ratings of the interventionalists regarding device functionality.Results: Eighty patients (39 women, mean age 74 ± 14 years) were eligible for our study. Median NIHSS at admission was 15 (IQR, 8-19), and median Alberta Stroke Program Early CT Score at baseline was 9 (IQR, 8-10). In 74 (93%) patients a primary combined approach was used as first-line technique. First-pass near-complete reperfusion was achieved in 43 (54%) and first-pass complete reperfusion in 34 (43%) patients. Final near-complete reperfusion was achieved in 66 (83%) patients after a median of 1.5 (1-3) passes, while final successful reperfusion was observed in 96% of our cases. We observed no device malfunctions. Median NIHSS at discharge was 2 (IQR, 0-5), and 3 patients (4%) suffered a symptomatic intracranial hemorrhage.Conclusions: Based on our initial data, we conclude that the Trevo NXT is an effective and safe tool for mechanical thrombectomy especially when used for combined approaches.
DOI:doi:10.3389/fneur.2021.704329
URL:Bitte beachten Sie: Dies ist ein Bibliographieeintrag. Ein Volltextzugriff für Mitglieder der Universität besteht hier nur, falls für die entsprechende Zeitschrift/den entsprechenden Sammelband ein Abonnement besteht oder es sich um einen OpenAccess-Titel handelt.

Volltext ; Verlag: https://doi.org/10.3389/fneur.2021.704329