Titel:Irreversible electroporation of the pig kidney with involvement of the renal pelvis
Titelzusatz:technical aspects, clinical outcome, and three-dimensional CT rendering for assessment of the treatment zone
Verf.angabe:Christof M. Sommer, Stefan Fritz, Miguel F. Wachter, Dominik Vollherbst, Ulrike Stampfl, Nadine Bellemann, Theresa Gockner, Theresa Mokry, Daniel Gnutzmann, Anne Schmitz, Jürgen Knapp, Thomas Longerich, Clemens Kuhn-Neureuther, Philippe L. Pereira, Hans U. Kauczor, Jens Werner, and Boris A. Radeleff
Jahr:2013
Umfang:10 S.
Fussnoten:Gesehen am 09.02.2022
Titel Quelle:Enthalten in: Journal of vascular and interventional radiology
Ort Quelle:Amsterdam [u.a.] : Elsevier, 1990
Jahr Quelle:2013
Band/Heft Quelle:24(2013), 12 vom: Dez., Seite 1888-1897
ISSN Quelle:1535-7732
Abstract:Purpose - To analyze irreversible electroporation (IRE) of the pig kidney with involvement of the renal pelvis. - Materials and Methods - IRE of renal tissue including the pelvis was performed in 10 kidneys in five pigs. Three study groups were defined: group I (two applicators with parallel configuration; n = 11), group II (three applicators with triangular configuration; n = 2), and group III (six applicators with complex configuration; n = 3). After IRE and before euthanasia, pigs underwent contrast-enhanced computed tomography (CT). Technical aspects (radial distance of applicators, resulting mean current), clinical outcome (complications, blood samples), and three-dimensional CT rendering for assessment of the treatment zone (short axis, circularity) were assessed. - Results - Radial distances of applicators were 14.3 mm ± 2.8 in group I, 12.3 mm ± 1.9 in group II, and 16.4 mm ± 3.5 in group III. Resulting mean currents were 25.7 A ± 6.5 in group I, 27.0 A ± 7.1 in group II, and 39.4 A ± 8.9 in group III. In group III, two perirenal hematomas were identified. There was no damage to the renal pelvis. During IRE, clinical blood parameters and cardiovascular markers did not change significantly. Short axis measurements were 20.6 mm ± 3.6 in group I, 31.9 mm ± 8.2 in group II, and 39.3 mm ± 2.4 in group III (P < .01 between groups). Circularity scores were 0.8 ± 0.2 in group I, 0.7 ± 0.1 in group II, and 0.7 ± 0.1 in group III, with a score of 1 indicating perfect roundness (P value not significant). - Conclusions - IRE of the pig kidney with involvement of the renal pelvis is feasible and safe. Size but not shape of the treatment zone is significantly affected by applicator configuration.
DOI:doi:10.1016/j.jvir.2013.08.014
URL:Bitte beachten Sie: Dies ist ein Bibliographieeintrag. Ein Volltextzugriff für Mitglieder der Universität besteht hier nur, falls für die entsprechende Zeitschrift/den entsprechenden Sammelband ein Abonnement besteht oder es sich um einen OpenAccess-Titel handelt.

Volltext ; Verlag: https://doi.org/10.1016/j.jvir.2013.08.014
 Volltext: https://www.sciencedirect.com/science/article/pii/S1051044313013833
 DOI: https://doi.org/10.1016/j.jvir.2013.08.014
Datenträger:Online-Ressource
Sprache:eng
K10plus-PPN:1789096626
Verknüpfungen:→ Zeitschrift

Permanenter Link auf diesen Titel (bookmarkfähig):  https://katalog.ub.uni-heidelberg.de/titel/68876375   QR-Code
zum Seitenanfang