Titel:Ultra-high dose rate (FLASH) carbon ion irradiation
Titelzusatz:dosimetry and first cell experiments
Verf.angabe:Walter Tinganelli, PhD, Olga Sokol, PhD, Martina Quartieri, MS, Anggraeini Puspitasari, MD, PhD, Ivana Dokic, PhD, Amir Abdollahi, MD, PhD, Marco Durante, PhD, Thomas Haberer, PhD, Jürgen Debus, MD, PhD, Daria Boscolo, PhD, Bernd Voss, PhD, Stephan Brons, PhD, Christoph Schuy, PhD, Felix Horst, PhD, and Ulrich Weber, PhD
Jahr:2022
Umfang:11 S.
Fussnoten:Available online 20 November 2021, Version of Record 18 February 2022 ; Gesehen am 28.03.2022
Titel Quelle:Enthalten in: International journal of radiation oncology, biology, physics
Ort Quelle:Amsterdam [u.a.] : Elsevier Science, 1975
Jahr Quelle:2022
Band/Heft Quelle:112(2022), 4, Seite 1012-1022
ISSN Quelle:1879-355X
Abstract:Purpose - To establish a beam monitoring and dosimetry system to enable the FLASH dose rate carbon ion irradiation and investigate, at different oxygen concentrations, the in vitro biological response in comparison to the conventional dose rate. - Methods and Materials - CHO-K1 cell response to irradiation at different dose rates and at different levels of oxygenation was studied using clonogenic assay. The Heidelberg Ion-Beam Therapy Center (HIT) synchrotron, after technical improvements, was adjusted to extract ≥5 × 108 12C ions within approximately 150 milliseconds. The beam monitors were filled with helium. - Results - The FLASH irradiation with beam scanning yields a dose of 7.5 Gy (homogeneity of ±5%) for a 280 MeV/u beam in a volume of at least 8 mm in diameter and a corresponding dose rate of 70 Gy/s (±20%). The dose repetition accuracy is better than 2%, the systematic uncertainty is better than 2%. Clonogenic assay demonstrates a significant FLASH sparing effect which is strongly oxygenation-dependent and mostly pronounced at 0.5% O2 but absent at 0% and 21% O2. - Conclusion - The FLASH dose rates >40 Gy/s were achieved with carbon beams. Cell survival analysis revealed FLASH dose rate sparing in hypoxia (0.5%-4% O2).
DOI:doi:10.1016/j.ijrobp.2021.11.020
URL:Bitte beachten Sie: Dies ist ein Bibliographieeintrag. Ein Volltextzugriff für Mitglieder der Universität besteht hier nur, falls für die entsprechende Zeitschrift/den entsprechenden Sammelband ein Abonnement besteht oder es sich um einen OpenAccess-Titel handelt.

Volltext: https://doi.org/10.1016/j.ijrobp.2021.11.020
 Volltext: https://www.sciencedirect.com/science/article/pii/S0360301621031291
 DOI: https://doi.org/10.1016/j.ijrobp.2021.11.020