Navigation überspringen
Universitätsbibliothek Heidelberg
Status: Bibliographieeintrag

Verfügbarkeit
Standort: ---
Exemplare: ---
heiBIB
 Online-Ressource
Verfasst von:Mayer, Christian [VerfasserIn]   i
 Son, Hyeck Soo [VerfasserIn]   i
 Łabuz, Grzegorz [VerfasserIn]   i
 Baur, Isabella [VerfasserIn]   i
 Yildirim, Timur Mert [VerfasserIn]   i
 Auffarth, Gerd U. [VerfasserIn]   i
 Khoramnia, Ramin [VerfasserIn]   i
Titel:Laboratory and clinical experience with a diffractive trifocal intraocular lens sutured to an artificial iris
Verf.angabe:Christian S. Mayer, MD; Hyeck-Soo Son, MD; Grzegorz Łabuz, PhD; Isabella D. Baur, MD; Timur M. Yildirim, MD; Gerd U. Auffarth, MD, PhD; Ramin Khoramnia, MD
E-Jahr:2022
Jahr:January 01, 2022
Umfang:11 S.
Fussnoten:Gesehen am 07.09.2022
Titel Quelle:Enthalten in: Journal of refractive surgery
Ort Quelle:Thorofare, NJ : SLACK Inc., 1996
Jahr Quelle:2022
Band/Heft Quelle:38(2022), 1, Seite 61-68
ISSN Quelle:1938-2391
Abstract:PURPOSE: - - To determine in vitro, using a translational research approach before realizing the procedure in a patient with iatrogenic aphakia and partial aniridia, whether suturing a trifocal intraocular lens (IOL) to an artificial iris degrades the IOL's optical quality. - - METHODS: - - Optical quality was analyzed by measuring the modulation transfer function (MTF) at a 3-mm aperture and at 50 and 100 lp/mm spatial frequencies. The FineVision Pod F GF IOL (PhysIOL) was assessed in two powers: two +20.00 diopters (D) (20A and 20B IOLs) and two +30.00 D (30A and 30B IOLs). The IOLs' decentration in relation to the artificial iris's center was evaluated. The laboratory results provided empirical evidence in the informed consent for surgical intervention in a patient with iatrogenic aphakia and iris defect in one eye. Clinical results were measured using the parameter of corrected distance visual acuity plus a patient self-assessment of the cosmetic appearance of the operated eye. - - RESULTS: - - The 20A and 20B IOLs demonstrated a mean MTF reduction of up to 1.1%, whereas the 30A and 30B IOLs showed a decrease of up to 5.2% for both spatial frequencies. All lenses showed good centration levels. In the clinical case, the patient showed corrected distance visual acuity, distance-corrected near visual acuity, and distance-corrected intermediate visual acuity of 0.20, 0.20, and 0.22 logMAR, respectively. The patient was satisfied with the cosmetic outcome. - - CONCLUSIONS: - - There was merely a slight reduction in trifocal IOL optical quality after it was sutured to an artificial iris. Clinically, the combined implantation of the artificial iris and FineVision IOL provided good functional and cosmetic outcomes. - - [J Refract Surg. 2022;38(1):61-68.]
DOI:doi:10.3928/1081597X-20211209-02
URL:Bitte beachten Sie: Dies ist ein Bibliographieeintrag. Ein Volltextzugriff für Mitglieder der Universität besteht hier nur, falls für die entsprechende Zeitschrift/den entsprechenden Sammelband ein Abonnement besteht oder es sich um einen OpenAccess-Titel handelt.

Volltext ; Verlag: https://doi.org/10.3928/1081597X-20211209-02
 Volltext: https://journals.healio.com/doi/10.3928/1081597X-20211209-02
 DOI: https://doi.org/10.3928/1081597X-20211209-02
Datenträger:Online-Ressource