Titel:Multicenter analytical evaluation of a high-sensitivity troponin T assay
Verf.angabe:A.K. Saenger, R. Beyrau, S. Braun, R. Cooray, A. Dolci, H. Freidank, E. Giannitsis, S. Gustafson, B. Handy, H. Katus, S.E. Melanson, M. Panteghini, P. Venge, M. Zorn, P. Jarolim, D. Bruton, J. Jarausch, A.S. Jaffe
E-Jahr:2011
Jahr:8 January 2011
Umfang:7 S.
Fussnoten:Gesehen am 30.09.2022
Titel Quelle:Enthalten in: Clinica chimica acta
Ort Quelle:Amsterdam [u.a.] : Elsevier Science, 1956
Jahr Quelle:2011
Band/Heft Quelle:412(2011), 9, Seite 748-754
ISSN Quelle:1873-3492
Abstract:Background - High-sensitivity cardiac troponin assays are being introduced clinically for earlier diagnosis of acute myocardial infarction (AMI). We evaluated the analytical performance of a high-sensitivity cardiac troponin T assay (hscTnT, Roche Diagnostics) in a multicenter, international trial. - Methods - Three US and 5 European sites evaluated hscTnT on the Modular® Analytics E170, cobas® 6000, Elecsys 2010, and cobas® e 411. Precision, accuracy, reportable range, an inter-laboratory comparison trial, and the 99th percentile of a reference population were assessed. - Results - Total imprecision (CVs) were 4.6-36.8% between 3.4 and 10.3ng/L hscTnT. Assay linearity was up to 10,000ng/L and the limit of blank and detection were 3 and 5ng/L, respectively. The 99th percentile reference limit was 14.2ng/L (n=533). No significant differences between specimen types, assay incubation time, or reagent lots existed. A substantial positive bias (76%) exists between the 4th generation and hscTnT assays at the low end of the measuring range (<50ng/L). hscTnT serum pool concentrations were within 2SD limits of the mean of means in the comparison trial, indicating comparable results across multiple platforms and laboratories. - Conclusion - The Roche hscTnT assay conforms to guideline precision requirements and will likely identify additional patients with myocardial injury suspicious for AMI.
DOI:doi:10.1016/j.cca.2010.12.034
URL:Bitte beachten Sie: Dies ist ein Bibliographieeintrag. Ein Volltextzugriff für Mitglieder der Universität besteht hier nur, falls für die entsprechende Zeitschrift/den entsprechenden Sammelband ein Abonnement besteht oder es sich um einen OpenAccess-Titel handelt.

Volltext ; Verlag: https://doi.org/10.1016/j.cca.2010.12.034
 Volltext: https://www.sciencedirect.com/science/article/pii/S0009898111000052
 DOI: https://doi.org/10.1016/j.cca.2010.12.034