Titel:Scheimpflug versus optical coherence tomography to detect subclinical corneal edema in Fuchs endothelial corneal dystrophy
Verf.angabe:Victor A. Augustin, Max K. Köppe, Hyeck-Soo Son, Jan Meis, Timur M. Yildirim, Ramin Khoramnia, Gerd U. Auffarth
E-Jahr:2022
Jahr:[November 2022]
Umfang:8 S.
Fussnoten:Gesehen am 04.01.2023
Titel Quelle:Enthalten in: Cornea
Ort Quelle:Philadelphia, Pa. : Lippincott, Williams & Wilkins, 1982
Jahr Quelle:2022
Band/Heft Quelle:41(2022), 11 vom: Nov., Seite 1378-1385
ISSN Quelle:1536-4798
Abstract:Purpose: - The purpose of this research was to compare the ability of Scheimpflug and anterior segment optical coherence tomography (OCT) in detecting subclinical corneal edema in patients with Fuchs endothelial corneal dystrophy (FECD) without clinical corneal edema. - Methods: - In this single-center, consecutive case series, 47 eyes of 29 patients with FECD were analyzed. The main outcome measures were anterior/posterior keratometry and central/thinnest corneal thickness. The criteria for subclinical corneal edema were loss of regular isopachs, displacement of the thinnest point of the cornea, and presence of posterior surface depression. Tomographic analyses were performed using Scheimpflug imaging (Pentacam HR) and OCT (anterior segment swept-source optical coherence tomography). - Results: - The measurement of the continuous variables revealed a significant difference between the 2 devices. The anterior curvature was steeper and the posterior curvature was flatter when measured with OCT (P < 0.001). The OCT showed a lower central corneal thickness and thinnest corneal thickness (P < 0.001). The agreement between both devices to detect subclinical corneal edema was high. The interdevice reliability for loss of parallel isopachs as measured by Cohen kappa coefficient was 0.84; for the displacement of the thinnest point of the cornea, it was 0.6, and for the focal posterior corneal surface depression, it was 0.62. Subclinical corneal edema was detected in 72.3% of the patients with both devices. In only 2 cases (4.3%), subclinical corneal edema was detected by one of the devices. - Conclusions: - Scheimpflug and OCT imaging were both able to detect tomographic patterns of subclinical corneal edema. Therefore, both devices can help decision-making, favoring early endothelial keratoplasty in symptomatic patients with FECD without clinical corneal edema.
DOI:doi:10.1097/ICO.0000000000002925
URL:Bitte beachten Sie: Dies ist ein Bibliographieeintrag. Ein Volltextzugriff für Mitglieder der Universität besteht hier nur, falls für die entsprechende Zeitschrift/den entsprechenden Sammelband ein Abonnement besteht oder es sich um einen OpenAccess-Titel handelt.

Volltext: https://doi.org/10.1097/ICO.0000000000002925
 Volltext: https://journals.lww.com/corneajrnl/Abstract/2022/11000/Scheimpflug_Versus_Optical_Coherence_Tomography_to.6.aspx