Navigation überspringen
Universitätsbibliothek Heidelberg
Standort: ---
Exemplare: ---
heiBIB
 Online-Ressource
Verfasst von:Klein, Valerie [VerfasserIn]   i
 Coll-Font, Jaume [VerfasserIn]   i
 Vendramini, Livia [VerfasserIn]   i
 Straney, Donald [VerfasserIn]   i
 Davids, Mathias [VerfasserIn]   i
 Ferris, Natalie G. [VerfasserIn]   i
 Schad, Lothar R. [VerfasserIn]   i
 Sosnovik, David E. [VerfasserIn]   i
 Nguyen, Christopher T. [VerfasserIn]   i
 Wald, Lawrence L. [VerfasserIn]   i
 Guérin, Bastien [VerfasserIn]   i
Titel:Measurement of magnetostimulation thresholds in the porcine heart
Verf.angabe:Valerie Klein, Jaume Coll-Font, Livia Vendramini, Donald Straney, Mathias Davids, Natalie G. Ferris, Lothar R. Schad, David E. Sosnovik, Christopher T. Nguyen, Lawrence L. Wald, Bastien Guérin
E-Jahr:2022
Jahr:30 July 2022
Umfang:17 S.
Fussnoten:Gesehen am 17.07.2023
Titel Quelle:Enthalten in: Magnetic resonance in medicine
Ort Quelle:New York, NY [u.a.] : Wiley-Liss, 1984
Jahr Quelle:2022
Band/Heft Quelle:88(2022), 5 vom: Nov., Seite 2242-2258
ISSN Quelle:1522-2594
Abstract:Purpose Powerful MRI gradient systems can surpass the International Electrotechnical Commission (IEC) 60601-2-33 limit for cardiac stimulation (CS), which was determined by simple electromagnetic simulations and electrode stimulation experiments. Only a few canine studies measured magnetically induced CS thresholds in vivo and extrapolating them to human safety limits can be challenging. Methods We measured cardiac magnetostimulation thresholds in 10 healthy, anesthetized pigs using capacitors discharged into a flat spiral coil to produce damped sinusoidal waveforms with effective stimulus duration ts,eff = 0.45 ms. Electrocardiography (ECG), blood pressure, and peripheral oximetry signals were recorded to determine threshold coil currents yielding cardiac capture. Dixon and CINE MR volumes from each animal were segmented to generate porcine-specific electromagnetic models to calculate dB/dt and E-field values in the porcine heart at threshold. For comparison, we also simulated maximum dB/dt and E-field values created by three MRI gradient systems in the heart of a human body model. Results The average dB/dt threshold estimated in the porcine heart was 1.66 ± 0.23 kT/s, which is 11-fold greater than the IEC dB/dt limit at ts,eff = 0.45 ms, and 31-fold greater than the maximum value created by the investigated MRI gradients in the human heart. The average E-field threshold estimated in the porcine heart was 92.9 ± 13.5 V/m, which is 6-fold greater than the IEC E-field limit at ts,eff = 0.45 ms and 37-fold greater than the maximum gradient-induced E-field in the human heart. Conclusion This first measurement of cardiac magnetostimulation thresholds in pigs indicates that the IEC cardiac safety limit is conservative for the investigated stimulus duration (ts,eff = 0.45 ms).
DOI:doi:10.1002/mrm.29382
URL:kostenfrei: Volltext: https://doi.org/10.1002/mrm.29382
 kostenfrei: Volltext: http://onlinelibrary.wiley.com/doi/abs/10.1002/mrm.29382
 DOI: https://doi.org/10.1002/mrm.29382
Datenträger:Online-Ressource
Sprache:eng
Sach-SW:cardiac stimulation
 electromagnetic exposure safety
 magnetostimulation thresholds
 MRI gradient field switching
K10plus-PPN:1852802731
Verknüpfungen:→ Zeitschrift
 
 
Lokale URL UB: Zum Volltext

Permanenter Link auf diesen Titel (bookmarkfähig):  https://katalog.ub.uni-heidelberg.de/titel/69098493   QR-Code