Titel:Optically stimulated luminescence detectors for dosimetry and LET measurements in light ion beams
Verf.angabe:Jeppe Brage Christensen, Iván Domingo Muñoz, Niels Bassler, Christina Stengl, Lily Bossin, Michele Togno, Sairos Safai, Oliver Jäkel and Eduardo Gardenali Yukihara
Jahr:2023
Umfang:15 S.
Fussnoten:Veröffentlicht: 21. Juli 2023 ; Gesehen am 29.08.2023
Titel Quelle:Enthalten in: Physics in medicine and biology
Ort Quelle:Bristol : IOP Publ., 1956
Jahr Quelle:2023
Band/Heft Quelle:68(2023), 15, Artikel-ID 155001, Seite 1-15
ISSN Quelle:1361-6560
Abstract:Objective. This work investigates the use of Al2O3:C and Al2O3:C,Mg optically stimulated luminescence (OSL) detectors to determine both the dose and the radiation quality in light ion beams. The radiation quality is here expressed through either the linear energy transfer (LET) or the closely related metric Q eff, which depends on the particle’s speed and effective charge. The derived LET and Q eff values are applied to improve the dosimetry in light ion beams. Approach. OSL detectors were irradiated in mono-energetic 1H-, 4He-, 12C-, and 16O-ion beams. The OSL signal is associated with two emission bands that were separated using a pulsed stimulation technique and subjected to automatic corrections based on reference irradiations. Each emission band was investigated independently for dosimetry, and the ratio of the two emission intensities was parameterized as a function of fluence- and dose-averaged LET, as well as Q eff. The determined radiation quality was subsequently applied to correct the dose for ionization quenching. Main results. For both materials, the Q eff determinations in 1H- and 4He-ion beams are within 5 % of the Monte Carlo simulated values. Using the determined radiation quality metrics to correct the nonlinear (ionization quenched) detector response leads to doses within 2 % of the reference doses. Significance. Al2O3:C and Al2O3:C,Mg OSL detectors are applicable for dosimetry and radiation quality estimations in 1H- and 4He-ions. Only Al2O3:C,Mg shows promising results for dosimetry in 12C-ions. Across both materials and the investigated ions, the estimated Q eff values were less sensitive to the ion types than the estimated LET values were. The reduced uncertainties suggest new possibilities for simultaneously estimating the physical and biological dose in particle therapy with OSL detectors.
DOI:doi:10.1088/1361-6560/acdfb0
URL:Bitte beachten Sie: Dies ist ein Bibliographieeintrag. Ein Volltextzugriff für Mitglieder der Universität besteht hier nur, falls für die entsprechende Zeitschrift/den entsprechenden Sammelband ein Abonnement besteht oder es sich um einen OpenAccess-Titel handelt.

kostenfrei: Volltext: https://doi.org/10.1088/1361-6560/acdfb0
 kostenfrei: Volltext: https://dx.doi.org/10.1088/1361-6560/acdfb0
 DOI: https://doi.org/10.1088/1361-6560/acdfb0
Datenträger:Online-Ressource
Sprache:eng
K10plus-PPN:1858137950
Verknüpfungen:→ Zeitschrift

Permanenter Link auf diesen Titel (bookmarkfähig):  https://katalog.ub.uni-heidelberg.de/titel/69116797   QR-Code
zum Seitenanfang