Titel: | Fast kilovoltage/megavoltage (kVMV) breathhold cone-beam CT for image-guided radiotherapy of lung cancer |
Verf.angabe: | Hansjoerg Wertz, Dzmitry Stsepankou, Manuel Blessing, Michael Rossi, Chris Knox, Kevin Brown, Uwe Gros, Judit Boda-Heggemann, Cornelia Walter, Juergen Hesser, Frank Lohr and Frederik Wenz |
E-Jahr: | 2010 |
Jahr: | 8 July 2010 |
Umfang: | 15 S. |
Illustrationen: | Illustrationen |
Fussnoten: | Gesehen am 29.11.2023 |
Titel Quelle: | Enthalten in: Physics in medicine and biology |
Ort Quelle: | Bristol : IOP Publ., 1956 |
Jahr Quelle: | 2010 |
Band/Heft Quelle: | 55(2010), 15, Seite 4203-4217 |
ISSN Quelle: | 1361-6560 |
Abstract: | Long image acquisition times of 60-120 s for cone-beam CT (CBCT) limit the number of patients with lung cancer who can undergo volume image guidance under breathhold. We developed a low-dose dual-energy kilovoltage-megavoltage-cone-beam CT (kVMV-CBCT) based on a clinical treatment unit reducing imaging time to ⩽15 s. Simultaneous kVMV-imaging was achieved by dedicated synchronization hardware controlling the output of the linear accelerator (linac) based on detector panel readout signals, preventing imaging artifacts from interference of the linac's MV-irradiation and panel readouts. Optimization was performed to minimize the imaging dose. Single MV-projections, reconstructed MV-CBCT images and images of simultaneous 90° kV- and 90° MV-CBCT (180° kVMV-CBCT) were acquired with different parameters. Image quality and imaging dose were evaluated and compared to kV-imaging. Hardware-based kVMV synchronization resulted in artifact-free projections. A combined 180° kVMV-CBCT scan with a total MV-dose of 5 monitor units was acquired in 15 s and with sufficient image quality. The resolution was 5-6 line pairs cm−1 (Catphan phantom). The combined kVMV-scan dose was equivalent to a kV-radiation scan dose of ∼33 mGy. kVMV-CBCT based on a standard linac is promising and can provide ultra-fast online volume image guidance with low imaging dose and sufficient image quality for fast and accurate patient positioning for patients with lung cancer under breathhold. |
DOI: | doi:10.1088/0031-9155/55/15/001 |
URL: | Bitte beachten Sie: Dies ist ein Bibliographieeintrag. Ein Volltextzugriff für Mitglieder der Universität besteht hier nur, falls für die entsprechende Zeitschrift/den entsprechenden Sammelband ein Abonnement besteht oder es sich um einen OpenAccess-Titel handelt.
Volltext: https://doi.org/10.1088/0031-9155/55/15/001 |
| Volltext: https://dx.doi.org/10.1088/0031-9155/55/15/001 |
| DOI: https://doi.org/10.1088/0031-9155/55/15/001 |
Datenträger: | Online-Ressource |
Sprache: | eng |
K10plus-PPN: | 1871582822 |
Verknüpfungen: | → Zeitschrift |
Fast kilovoltage/megavoltage (kVMV) breathhold cone-beam CT for image-guided radiotherapy of lung cancer / Wertz, Hansjörg [VerfasserIn]; 8 July 2010 (Online-Ressource)