Navigation überspringen
Universitätsbibliothek Heidelberg
Status: Bibliographieeintrag

Verfügbarkeit
Standort: ---
Exemplare: ---
heiBIB
 Online-Ressource
Verfasst von:Khoramnia, Ramin [VerfasserIn]   i
 Baur, Isabella [VerfasserIn]   i
 Łabuz, Grzegorz [VerfasserIn]   i
 Chychko, Lizaveta [VerfasserIn]   i
 Köppe, Maximilian [VerfasserIn]   i
 Hallak, Mustafa [VerfasserIn]   i
 Auffarth, Gerd U. [VerfasserIn]   i
Titel:Enlargement of main corneal incision
Titelzusatz:clinical intraindividual comparison of two preloaded intraocular lens injectors
Verf.angabe:Ramin Khoramnia, MD, PhD, FEBO, Isabella D. Baur, MD, Grzegorz Łabuz, PhD, Lizaveta Chychko, Maximilian K. Köppe, MD, Mustafa K. Hallak, MD, Gerd U. Auffarth, MD, PhD, FEBO
E-Jahr:2023
Jahr:February 2023
Umfang:6 S.
Illustrationen:Illustrationen
Fussnoten:Gesehen am 14.12.2023
Titel Quelle:Enthalten in: Journal of cataract and refractive surgery
Ort Quelle:Fairfax, Va. : Soc., 1986
Jahr Quelle:2023
Band/Heft Quelle:49(2023), 2 vom: Feb., Artikel-ID 165, Seite 1-6
ISSN Quelle:1873-4502
Abstract:Purpose: - To compare the enlargement of the clear corneal incision from IOL implantation with 2 different intraocular lens (IOL) injectors: the AutonoMe preloaded with the Clareon IOL and the Multisert preloaded with the Vivinex IOL. - Setting: - The David J. Apple Center for Vision Research, Department of Ophthalmology, University of Heidelberg, Heidelberg, Germany. - Design: - Prospective randomized clinical comparative study. - Methods: - 96 eyes of 48 patients with cataract were intraindividually randomized to treatment with 1 of the 2 injectors. For Multisert eyes, the insert shield (IS) was used in the advanced position in 23 eyes. The initial incision was 2.2 mm, and intraoperative measurements of the incision size were made before and after IOL injection. 3 months postoperatively, keratometry and uncorrected (UDVA) and corrected (CDVA) distance visual acuities were assessed. - Results: - Results are reported for 96 eyes of 48 patients. The mean incision enlargement was 0.213 ± 0.068 mm in the Multisert with the IS group, 0.265 ± 0.055 mm in the fellow eyes (AutonoMe) (P < .05), 0.272 ± 0.060 mm in Multisert eyes treated without the IS, and 0.296 ± 0.066 mm for the fellow eyes (AutonoMe) (P > .05). The mean absolute surgically induced astigmatism was 0.42 ± 0.23 diopters (D), 0.50 ± 0.25 D, and 0.44 ± 0.18 D in the Multisert with the IS, Multisert without the IS, and AutonoMe group, respectively (P > .05). The UDVA and CDVA were comparable in all groups. - Conclusions: - The Multisert was associated with less wound enlargement than the AutonoMe. All groups had comparable functional outcomes. Therefore, the observed difference in incision enlargement may be of limited clinical relevance.
DOI:doi:10.1097/j.jcrs.0000000000001064
URL:Bitte beachten Sie: Dies ist ein Bibliographieeintrag. Ein Volltextzugriff für Mitglieder der Universität besteht hier nur, falls für die entsprechende Zeitschrift/den entsprechenden Sammelband ein Abonnement besteht oder es sich um einen OpenAccess-Titel handelt.

Volltext: https://doi.org/10.1097/j.jcrs.0000000000001064
 Volltext: https://journals.lww.com/jcrs/fulltext/2023/02000/enlargement_of_main_corneal_incision__clinical.10.aspx
 DOI: https://doi.org/10.1097/j.jcrs.0000000000001064
Datenträger:Online-Ressource
Sprache:eng
K10plus-PPN:1875843973
Verknüpfungen:→ Zeitschrift

Permanenter Link auf diesen Titel (bookmarkfähig):  https://katalog.ub.uni-heidelberg.de/titel/69152835   QR-Code
zum Seitenanfang