Titel:In vitro comparison of two electromagnetic shockwave generators: low-pressure wide focus vs high-pressure small focus
Titelzusatz:impact on initial stone fragmentation and final stone comminution
Verf.angabe:Jens Rassweiler, Philip Rieker, Rainer Pecha, Martin Dressel, and Marie-Claire Rassweiler-Seyfried
E-Jahr:2022
Jahr:1 February 2022
Umfang:7 S.
Fussnoten:Gesehen am 24.04.2024
Titel Quelle:Enthalten in: Journal of endourology
Ort Quelle:Larchmont, NY : Liebert, 1999
Jahr Quelle:2022
Band/Heft Quelle:36(2022), 2, Seite 266-272
ISSN Quelle:1557-900X
Abstract:Context: Recently developed concepts for higher efficacy extracorporeal shockwave lithotripsy with low-pressure wide focus systems resulting in finer fragmentation of the calculi. - Objective: To compare two different electromagnetic shockwave sources (low-pressure wide focus [Xinin Lithotripter, XL] vs high-pressure small focus [Siemens Lithoskop, SL]) by sound field measurements and in vitro fragmentation. - Evidence Acquisition: The CS-2012A XX-ES lithotripter (self-focusing electromagnetic shockwave generator with concave spherical curved electrical coil; XL) was compared to the (SL) (electromagnetic generator with a flat electric coil with an acoustical lens). Different sound field measurements were performed using a fiber-optic hydrophone. Measurements at three different power settings (XL: 8.0, 9.3, and 10.3 kV and SL: Level 1, 5, and 8). Ten AST stones and 15 BegoStones (9.3 kV, Level 3) with a frequency of 90/minute (SL) and 20/minute (XL). Number of impulses to the first crack and for complete stone comminution (residual fragments <2 mm) was documented. - Results: The median number of shockwaves for the first crack in AST stones with the XL was 12 (10-14) and 7 with the SL (6-9). Complete disintegration was accomplished after 815 (782-824) shockwaves with XL and 702 (688-712) with SL. The difference was not statistically significant. The median number of shockwaves to produce the first crack in BegoStones was 524 (504-542) with XL and only 151 (137-161) with SL. Numbers of shockwaves for complete disintegration did not differ significantly (XL: 2518 vs SL: 2287). Using a wide focus with low pressure shows more homogeneous disintegration. - Conclusion: Two stone models showed significant differences regarding form and time of the initial fragmentation. Impulses for stone comminution did not differ significantly. The advantages of a low-pressure wide-focus system include minimal trauma and a homogeneous fragment size but is more time consuming. High-pressure small-focus systems are clinically effective.
DOI:doi:10.1089/end.2021.0416
URL:Bitte beachten Sie: Dies ist ein Bibliographieeintrag. Ein Volltextzugriff für Mitglieder der Universität besteht hier nur, falls für die entsprechende Zeitschrift/den entsprechenden Sammelband ein Abonnement besteht oder es sich um einen OpenAccess-Titel handelt.

Volltext: https://doi.org/10.1089/end.2021.0416
 Volltext: http://www.liebertpub.com/doi/10.1089/end.2021.0416
 DOI: https://doi.org/10.1089/end.2021.0416
Datenträger:Online-Ressource
Sprache:eng
K10plus-PPN:1888537892
Verknüpfungen:→ Zeitschrift

Permanenter Link auf diesen Titel (bookmarkfähig):  https://katalog.ub.uni-heidelberg.de/titel/69213585   QR-Code
zum Seitenanfang