Titel:Evaluation of parameters and nozzle tip damage after clinical use of three hydrophilic intraocular lens injector models
Verf.angabe:Lu Zhang, Sonja Schickhardt, Patrick Merz, and Gerd Uwe Auffarth
E-Jahr:2024
Jahr:30 May 2024
Umfang:8 S.
Illustrationen:Illustrationen
Fussnoten:Gesehen am 10.10.2024
Titel Quelle:Enthalten in: Journal of ophthalmology
Ort Quelle:New York, NY : Hindawi, 2009
Jahr Quelle:2024
Band/Heft Quelle:(2024), Artikel-ID 2360368, Seite 1-8
ISSN Quelle:2090-0058
Abstract:Purpose. To assess the nozzle tip damage and the parameters of three different hydrophilic intraocular lens (IOL) injector models. Methods. After routine cataract surgeries at the University Eye Hospital Heidelberg, all the used IOL injectors were collected from the operating room and sent to our laboratory. Nozzle tip damage was assessed under a microscope and graded as follows: no damage (grade 0), slight scratches (1), deep scratches (2), extensions (3), cracks (4), and bursts (5). Each damage grade was assigned a score from 0 to 5, and the total damage score for each injector system was calculated and compared. Nozzle tip parameters (diameters and areas), plunger tip parameters, and tip angles were also measured in each model. Results. The damage scores were (median, Q3-Q1): 1 (1-1) for Accuject, 1 (1-1) for Bluemixs, and 1 (1-1) for RayOne. There was no statistically significant difference in the damage scores between the study groups (P > 0.05). The outer cross-sectional vertical and horizontal diameters were 1.69 and 1.69 mm for Accuject, 1.69 and 1.69 mm for Bluemixs, and 1.70 and 1.71 mm for RayOne. Plunger tip areas were 0.78 mm2 for Accjuect, 0.74 mm2 for Bluemixs, and 0.43 mm2 for RayOne. Plunger tip area/inner cross-sectional area of the nozzle tip (%) was 31.2% for RayOne, 66.7% for Accuject, and 63.8% for Bluemixs. The tip angles for three injector models were 56° (Accuject), 56° (Bluemixs), and 44° (RayOne). Conclusions. All the injector models showed mild to moderate damage to the nozzle tip after IOL implantation, even with smaller diameter tips. RayOne resulted in the lowest ratio between plunger tip area and inner cross-sectional area of the nozzle tip and a better distribution of damage categories than the other two groups. All three injector models had relatively small tip parameters. If smaller incisions are required in certain patients, smaller tip parameters should be considered.
DOI:doi:10.1155/2024/2360368
URL:Bitte beachten Sie: Dies ist ein Bibliographieeintrag. Ein Volltextzugriff für Mitglieder der Universität besteht hier nur, falls für die entsprechende Zeitschrift/den entsprechenden Sammelband ein Abonnement besteht oder es sich um einen OpenAccess-Titel handelt.

kostenfrei: Volltext: https://doi.org/10.1155/2024/2360368
 kostenfrei: Volltext: https://onlinelibrary.wiley.com/doi/abs/10.1155/2024/2360368
 DOI: https://doi.org/10.1155/2024/2360368
Datenträger:Online-Ressource
Sprache:eng
K10plus-PPN:1905362595
Verknüpfungen:→ Zeitschrift

Permanenter Link auf diesen Titel (bookmarkfähig):  https://katalog.ub.uni-heidelberg.de/titel/69261306   QR-Code
zum Seitenanfang