Titel:A body mass index-based method for “MR-only” abdominal MR-guided adaptive radiotherapy
Verf.angabe:Carolin Rippke, C. Katharina Renkamp, Christiane Stahl-Arnsberger, Annette Miltner, Carolin Buchele, Juliane Hörner-Rieber, Jonas Ristau, Jürgen Debus, Markus Alber, Sebastian Klüter
E-Jahr:2024
Jahr:August 2024
Umfang:12 S.
Illustrationen:Illustrationen
Fussnoten:Online verfügbar: 8. Februar 2023, Artikelversion: 14. August 2024 ; Gesehen am 11.10.2024
Titel Quelle:Enthalten in: Zeitschrift für medizinische Physik
Ort Quelle:Amsterdam [u.a.] : Elsevier, 1990
Jahr Quelle:2024
Band/Heft Quelle:34(2024), 3 vom: Aug., Seite 456-467
ISSN Quelle:1876-4436
Abstract:Purpose - Dose calculation for MR-guided radiotherapy (MRgRT) at the 0.35 T MR-Linac is currently based on deformation of planning CTs (defCT) acquired for each patient. We present a simple and robust bulk density overwrite synthetic CT (sCT) method for abdominal treatments in order to streamline clinical workflows. - Method - Fifty-six abdominal patient treatment plans were retrospectively evaluated. All patients had been treated at the MR-Linac using MR datasets for treatment planning and plan adaption and defCT for dose calculation. Bulk density CTs (4M-sCT) were generated from MR images with four material compartments (bone, lung, air, soft tissue). The relative electron densities (RED) for bone and lung were extracted from contoured CT structure average REDs. For soft tissue, a correlation between BMI and RED was evaluated. Dose was recalculated on 4M-sCT and compared to dose distributions on defCTs assessing dose differences in the PTV and organs at risk (OAR). - Results - Mean RED of bone was 1.17±0.02, mean RED of lung 0.17±0.05. The correlation between BMI and RED for soft tissue was statistically significant (p<0.01). PTV dose differences between 4M-sCT and defCT were Dmean: −0.4±1.0%, D1%: −0.3±1.1% and D95%: −0.5±1.0%. OARs showed D2%: −0.3±1.9% and Dmean: −0.1±1.4% differences. Local 3D gamma index pass rates (2%/2mm) between dose calculated using 4M-sCT and defCT were 96.8±2.6% (range 89.9-99.6%). - Conclusion - The presented method for sCT generation enables precise dose calculation for MR-only abdominal MRgRT.
DOI:doi:10.1016/j.zemedi.2022.12.001
URL:Bitte beachten Sie: Dies ist ein Bibliographieeintrag. Ein Volltextzugriff für Mitglieder der Universität besteht hier nur, falls für die entsprechende Zeitschrift/den entsprechenden Sammelband ein Abonnement besteht oder es sich um einen OpenAccess-Titel handelt.

kostenfrei: Volltext: https://doi.org/10.1016/j.zemedi.2022.12.001
 kostenfrei: Volltext: https://www.sciencedirect.com/science/article/pii/S0939388922001349
 DOI: https://doi.org/10.1016/j.zemedi.2022.12.001
Datenträger:Online-Ressource
Sprache:eng
Sach-SW:Adaptive radiotherapy
 Bulk density
 Image guided radiotherapy
 MR guided radiotherapy
 MR-only
 Synthetic CT
K10plus-PPN:1905492774
Verknüpfungen:→ Zeitschrift

Permanenter Link auf diesen Titel (bookmarkfähig):  https://katalog.ub.uni-heidelberg.de/titel/69261615   QR-Code
zum Seitenanfang