Titel:Cerebral arteriovenous malformations
Titelzusatz:Improved nidus demarcation by means of dynamic tagging MR-angiography
Verf.angabe:M. Essig, R. Engenhart, M.V. Knopp, M. Bock, J. Scharf, J. Debus, F. Wenz, H. Hawighorst, L.R. Schad, G. van Kaick
Jahr:1996
Umfang:7 S.
Fussnoten:Elektronische Reproduktion der Druck-Ausgabe 8. Dezember 1999 ; Gesehen am 12.12.2024
Titel Quelle:Enthalten in: Magnetic resonance imaging
Ort Quelle:Amsterdam [u.a.] : Elsevier Science, 1982
Jahr Quelle:1996
Band/Heft Quelle:14(1996), 3, Seite 227-233
ISSN Quelle:1873-5894
Abstract:Our purpose was to further improve the target volume definition for radiosurgical treatment of cerebral arteriovenous malformations (AVMs) by means of dynamic MRA (dMRA) using a blood bolus tagging sequence. We therefore compare this technique with 3D-TOF-MRA and transfemoral high resolution angiography in plain film technique. Twenty patients with angiographically proven cerebral AVMs were investigated by dMRA, TOF-MRA, and conventional angiography during the MR-assisted radiosurgical planning protocol. The patient's head was fixed in an MR-compatible stereotactic device. The different angiography techniques were evaluated by consensus of two radiologists. AVMs were characterized by the number and origin of feeding arteries, the maximum diameter of the AVM nidus, and the venous drainage pattern. Dynamic MRA was able to demonstrate the complete AVM characteristics and hemodynamics in 12 out of 20 patients. In three patients with an AVM nidus smaller than 1 cm in diameter, the technique could not reliably depict the malformation. Technical problems due to steel skrews and pins in the initially used stereotactic frame occured in five patients. Due to reduced vessel overlap and the lack of disturbances caused by formations with short T1 time, dMRA was superior to TOF-MRA in the detection and the exact localization of the AVM nidus in four patients. We conclude that dMRA is able to demonstrate reliably AVM characteristics and hemodynamics in AVMs with a nidus larger than 1 cm in diameter. Because of the improved demarcation of the AVM nidus, this technique may be a valuable adjunct to radiosurgery planning of cerebral AVMs.
DOI:doi:10.1016/0730-725X(95)02102-Y
URL:Bitte beachten Sie: Dies ist ein Bibliographieeintrag. Ein Volltextzugriff für Mitglieder der Universität besteht hier nur, falls für die entsprechende Zeitschrift/den entsprechenden Sammelband ein Abonnement besteht oder es sich um einen OpenAccess-Titel handelt.

Volltext: https://doi.org/10.1016/0730-725X(95)02102-Y
 Volltext: https://www.sciencedirect.com/science/article/pii/0730725X9502102Y
 DOI: https://doi.org/10.1016/0730-725X(95)02102-Y
Datenträger:Online-Ressource
Sprache:eng
Sach-SW:AVM
 Blood bolus tagging
 MR-Angiography
 Radiosurgery planning
K10plus-PPN:1912171074
Verknüpfungen:→ Zeitschrift

Permanenter Link auf diesen Titel (bookmarkfähig):  https://katalog.ub.uni-heidelberg.de/titel/69283202   QR-Code
zum Seitenanfang