Titel:Spectral performance evaluation of a second-generation spectral detector CT
Verf.angabe:Leening P. Liu, Nadav Shapira, Sandra S. Halliburton, Sebastian Meyer, Amy Perkins, Harold I. Litt, Hans Ulrich Kauczor, Tim Leiner, Wolfram Stiller, Peter B. Noël
E-Jahr:2024
Jahr:April 2024
Umfang:13 S.
Illustrationen:Illustrationen
Fussnoten:Gesehen am 18.02.2025
Titel Quelle:Enthalten in: Journal of applied clinical medical physics
Ort Quelle:Reston, Va. : ACMP, 2000
Jahr Quelle:2024
Band/Heft Quelle:25(2024), 4 vom: Apr., Artikel-ID e14300, Seite e14300-1-e14300-13
ISSN Quelle:1526-9914
Abstract:Purpose The aim of this study was to characterize a second-generation wide-detector dual-layer spectral computed tomography (CT) system for material quantification accuracy, acquisition parameter and patient size dependencies, and tissue characterization capabilities. Methods A phantom with multiple tissue-mimicking and material-specific inserts was scanned with a dual-layer spectral detector CT using different tube voltages, collimation widths, radiation dose levels, and size configurations. Accuracy of iodine density maps and virtual monoenergetic images (MonoE) were investigated. Additionally, differences between conventional and MonoE 70 keV images were calculated to evaluate acquisition parameter and patient size dependencies. To demonstrate material quantification and differentiation, liver-mimicking inserts with adipose and iron were analyzed with a two-base decomposition utilizing MonoE 50 and 150 keV, and root mean square error (RMSE) for adipose and iron content was reported. Results Measured inserts exhibited quantitative accuracy across a wide range of MonoE levels. MonoE 70 keV images demonstrated reduced dependence compared to conventional images for phantom size (1 vs. 27 HU) and acquisition parameters, particularly tube voltage (4 vs. 37 HU). Iodine density quantification was successful with errors ranging from -0.58 to 0.44 mg/mL. Similarly, inserts with different amounts of adipose and iron were differentiated, and the small deviation in values within inserts corresponded to a RMSE of 3.49 ± 1.76% and 1.67 ± 0.84 mg/mL for adipose and iron content, respectively. Conclusion The second-generation dual-layer CT enables acquisition of quantitatively accurate spectral data without compromises from differences in patient size and acquisition parameters.
DOI:doi:10.1002/acm2.14300
URL:Bitte beachten Sie: Dies ist ein Bibliographieeintrag. Ein Volltextzugriff für Mitglieder der Universität besteht hier nur, falls für die entsprechende Zeitschrift/den entsprechenden Sammelband ein Abonnement besteht oder es sich um einen OpenAccess-Titel handelt.

kostenfrei: Volltext: https://doi.org/10.1002/acm2.14300
 kostenfrei: Volltext: https://onlinelibrary.wiley.com/doi/abs/10.1002/acm2.14300
 DOI: https://doi.org/10.1002/acm2.14300