Titel:Comparison of surgical times between manual and robot-assisted epiretinal membrane peeling
Verf.angabe:Alexander Eberle, Ferhat Turgut, Gábor Márk Somfai, Amr Saad, Marc D. de Smet, Pascal W. Hasler, Florian M. Heussen, and Matthias D. Becker
E-Jahr:2024
Jahr:August 14, 2024
Umfang:7 S.
Illustrationen:Illustrationen
Fussnoten:Gesehen am 07.03.2025
Titel Quelle:Enthalten in: Translational Vision Science & Technology
Ort Quelle:Rockville, Md. : ARVO, 2012
Jahr Quelle:2024
Band/Heft Quelle:13(2024), 8, Artikel-ID 27, Seite 1-7
ISSN Quelle:2164-2591
Abstract:Epiretinal membranes (ERM) pose a common challenge in vitreoretinal pathology, often causing vision impairment in older adults. The Preceyes Surgical System (PSS) supports the surgical removal of ERM through robot-assisted membrane peeling (RA-MP). This study compares surgical times and iatrogenic hemorrhages between manual membrane peeling (MMP) and RA-MP using PSS. Nine patients underwent RA-MP with PSS, whereas 16 patients (18 eyes) underwent MMP for comparative analysis. Surgical durations were categorized into RA-MP, manual forceps utilization in PSS surgeries (mRA-MP), and traditional MMP. Cumulative manual manipulation duration (cMMP), instrument grasps, and intraoperative hemorrhages were statistically analyzed using the Mann-Whitney U test. RA-MP showed significantly longer peeling times compared to MMP (P < 0.001). Flap initiation grasps were similar between methods (P = 0.86), RA-MP demonstrated a significant reduction in peeling grasps (P = 0.01) and mean grasps per minute (P < 0.001). Although RA-MP resulted in fewer hemorrhages, the difference did not reach statistical significance relative to MMP (P = 0.08). Although RA-MP tended to extend surgical time, it offered advantages in reducing tissue trauma and intraoperative hemorrhages. Further research is needed to explore the learning curve for novice surgeons and evaluate the safety profile of RA-MP. RA-MP may offer potential advantages over manual surgery, particularly in terms of reduced tissue trauma and intraoperative hemorrhages. Despite its longer duration compared with manual techniques, RA-MP may lead to fewer grasping maneuvers and lower rates of hemorrhages, thereby enhancing the safety and precision of vitreoretinal surgeries.
DOI:doi:10.1167/tvst.13.8.27
URL:Bitte beachten Sie: Dies ist ein Bibliographieeintrag. Ein Volltextzugriff für Mitglieder der Universität besteht hier nur, falls für die entsprechende Zeitschrift/den entsprechenden Sammelband ein Abonnement besteht oder es sich um einen OpenAccess-Titel handelt.

kostenfrei: Volltext: https://doi.org/10.1167/tvst.13.8.27
 DOI: https://doi.org/10.1167/tvst.13.8.27
Datenträger:Online-Ressource
Sprache:eng
K10plus-PPN:1919306137
Verknüpfungen:→ Zeitschrift

Permanenter Link auf diesen Titel (bookmarkfähig):  https://katalog.ub.uni-heidelberg.de/titel/69316097   QR-Code
zum Seitenanfang