Titel:CT volumetry of artificial pulmonary nodules using an ex vivo lung phantom
Titelzusatz:influence of exposure parameters and iterative reconstruction on reproducibility
Verf.angabe:Mark O. Wielpütz, Mathieu Lederlin, Jacek Wroblewski, Julien Dinkel, Monika Eichinger, Jürgen Biederer, Hans-Ulrich Kauczor, Michael Puderbach
E-Jahr:2013
Jahr:31 May 2013
Umfang:7 S.
Fussnoten:Gesehen am 08.11.2021
Titel Quelle:Enthalten in: European journal of radiology
Ort Quelle:Amsterdam [u.a.] : Elsevier Science, 1990
Jahr Quelle:2013
Band/Heft Quelle:82(2013), 9, Seite 1577-1583
ISSN Quelle:1872-7727
Abstract:Objectives - To evaluate the influence of exposure parameters and raw-data based iterative reconstruction (IR) on the measurement variability of computer-aided nodule volumetry on chest multidetector computed tomography (MDCT). - Materials and methods - N=7 porcine lung explants were inflated in a dedicated ex vivo phantom and prepared with n=162 artificial nodules. MDCT was performed eight consecutive times (combinations of 120 and 80kV with 120, 60, 30 and 12mAs), and reconstructed with filtered back projection (FBP) and IR. Nodule volume and diameter were measured semi-automatically with dedicated software. The absolute percentage measurement error (APE) was computed in relation to the 120kV 120mAs acquisition. Noise was recorded for each nodule in every dataset. - Results - Mean nodule volume and diameter were 0.32±0.15ml and 12.0±2.6mm, respectively. Although IR reduced noise by 24.9% on average compared to FBP (p<0.007), APE with IR was equal to or slightly higher than with FBP. Mean APE for volume increased significantly below a volume computed tomography dose index (CTDI) of 1.0mGy: for 120kV 12mAs APE was 3.8±6.2% (FBP) vs. 4.0±5.2% (IR) (p<0.007); for 80kV 12mAs APE was 8.0±13.0% vs. 9.3±15.8% (n.s.), respectively. Correlating APE with image noise revealed that at identical noise APE was higher with IR than with FBP (p<0.05). - Conclusions - Computer-aided volumetry is robust in a wide range of exposure settings, and reproducibility is reduced at a CTDI below 1.0mGy only, but the error rate remains clinically irrelevant. Noise reduction by IR is not detrimental for measurement error in the setting of semi-automatic nodule volumetry on chest MDCT.
DOI:doi:10.1016/j.ejrad.2013.04.035
URL:Bitte beachten Sie: Dies ist ein Bibliographieeintrag. Ein Volltextzugriff für Mitglieder der Universität besteht hier nur, falls für die entsprechende Zeitschrift/den entsprechenden Sammelband ein Abonnement besteht oder es sich um einen OpenAccess-Titel handelt.

Volltext ; Verlag: https://doi.org/10.1016/j.ejrad.2013.04.035
 Volltext: https://www.sciencedirect.com/science/article/pii/S0720048X13002325
 DOI: https://doi.org/10.1016/j.ejrad.2013.04.035
Datenträger:Online-Ressource
Sprache:eng
Sach-SW:Computer-aided diagnosis
 Lung cancer
 Screening
K10plus-PPN:1776305434
Verknüpfungen:→ Zeitschrift

Permanenter Link auf diesen Titel (bookmarkfähig):  https://katalog.ub.uni-heidelberg.de/titel/68797673   QR-Code
zum Seitenanfang