Titel: | A comparative analysis of the effects of misaligning different trifocal intraocular lenses |
Verf.angabe: | Weijia Yan, Gerd U. Auffarth, Ramin Khoramnia and Grzegorz Łabuz |
E-Jahr: | 2025 |
Jahr: | January 2025 |
Umfang: | 11 S. |
Illustrationen: | Illustrationen |
Fussnoten: | Gesehen am 18.06.2025 |
Titel Quelle: | Enthalten in: Journal of Clinical Medicine |
Ort Quelle: | Basel : MDPI, 2012 |
Jahr Quelle: | 2025 |
Band/Heft Quelle: | 14(2025), 1, Artikel-ID 187, Seite 1-11 |
ISSN Quelle: | 2077-0383 |
Abstract: | Objectives: This laboratory study aims to assess the effects of misaligning different trifocal intraocular lenses (IOLs) under varying spectral and corneal spherical aberration (SA) conditions. Methods: With an IOL metrology device under monochromatic and polychromatic conditions, the following models were studied: AT ELANA 841P, AT LISA Tri 839MP, FineVision HP POD F, Acrysof IQ PanOptix, and Tecnis Synergy ZFR00V. The SA was simulated using an aberration-free and average-SA cornea. The modulation transfer function (MTF) was measured at different pupil sizes for the on- and off-axis lens positions. Results: The IOLs exhibited varying responses to decentration up to 1 mm, showing the lowest impact in polychromatic light. The least affected was AT ELANA, with an MTF loss of 15.7% to 28.4% at 50 lp/mm across the studied conditions. It was followed by PanOptix and FineVision, with the MTF loss ranging from 19.1% to 36.0% and from 21.2% to 46.6%. AT LISA showed a more substantial reduction, i.e., 41.2% to 64.8%, but it was still lower than that of Synergy (51.1% to 78.8%). When decentration was induced at a 4.5 mm distance, its effect was more evident in conditions that were closer to each IOL’s SA correction. A tilt of 5° had a lesser impact than 1 mm decentration, with the effect being more severe at 4.5 mm. Conclusions: The off-axis position affects the optical quality of trifocal IOLs. Low- rather than high-SA-correcting trifocals perform better under misalignment. In polychromatic light, the impact of misalignment is less evident, suggesting a potential mitigating effect of chromatic aberration. |
DOI: | doi:10.3390/jcm14010187 |
URL: | Bitte beachten Sie: Dies ist ein Bibliographieeintrag. Ein Volltextzugriff für Mitglieder der Universität besteht hier nur, falls für die entsprechende Zeitschrift/den entsprechenden Sammelband ein Abonnement besteht oder es sich um einen OpenAccess-Titel handelt.
kostenfrei: Volltext: https://doi.org/10.3390/jcm14010187 |
| kostenfrei: Volltext: https://www.mdpi.com/2077-0383/14/1/187 |
| DOI: https://doi.org/10.3390/jcm14010187 |
Datenträger: | Online-Ressource |
Sprache: | eng |
Sach-SW: | decentration |
| modulation transfer function |
| optical quality |
| tilt |
| trifocal IOLs |
K10plus-PPN: | 1928518680 |
Verknüpfungen: | → Zeitschrift |
¬A¬ comparative analysis of the effects of misaligning different trifocal intraocular lenses / Yan, Weijia [VerfasserIn]; January 2025 (Online-Ressource)
69357350